IQ Analyzers and Interface Options0 pages
THERMOX
®
PROCESS INSTRUMENTS
IQ™ Analyzers and Interface Options
IQ ANALYZERS
The IQ sensor includes the necessary
intelligence for proper operation
directly on the sensor eliminating the
need for a separate control unit. The
electronics contained in the sensor
include auto calibration, analog
output, alarms, self-diagnostics and
network communications.
The IQ or smart sensor line provides
local and field communications.
The new design allows access to
process variable and diagnostic
data previously inaccessible using
conventional analog outputs. IQ
analyzers are available with HART®
or RS-485 AMETEK protocol
communications options. User
interface with the IQ sensor is
achieved through a HART-compatible
device, IQ Link Terminal or Series
2000 Host.
FEATURES
►t Electronics contained in the
sensor means no interconnecting
cable and no control unit making
installation and operation easier
and less expensive.
►t Automatic calibration (with
optional remote calibration unit),
analog outputs, alarms, selfdiagnostics and network
communications
►t Wall-mount or handheld IQ Links
allow set up or display of sensor
data either locally or remotely.
►t HART® option available. IQ
analyzers work with other
HART communicators or other
HART protocol systems. Device
definitions registered with HART
foundation.
►t Series 2000 Host Controller
option for averaging up to 4
channels, or setup of up to 32 IQ
sensors
►t Retrofit existing analyzers
►t Upgrade to network serial
communications
Handheld IQ Link
Wall Mount IQ Link
IQ™ LINKS
Handheld or dedicated wall mount
IQ Links provide a local interface to
the Thermox line of Smart Sensors
using RS-232 communications.
The IQ links can be used to set up
and display system parameters,
initiate calibrations, and perform
system tests and troubleshooting.
The handheld link can be plugged
into any Thermox IQ Analyzer. The
dedicated wall mount version is
hard wired to a specific analyzer
but can be rewired depending on
plant needs. Both the hand-held
and wall mount terminals are
approved for Class I, Division 2,
Groups A, B, C and D.
t The handheld and dedicated wall
mount IQ Links work in cooperation
with the Series 2000 Host which
can be used to set up to 32
analyzers via an RS-485 bus.
Interface:
Handheld: 7 ft. (213 cm) plug-in
RS-232
Wall Mount: RS-232, max. distance
50 ft. (1524 cm)
Display:
Four-line by 20-character extended
temperature Supertwist Nematic
LCD.
Enclosure:
Handheld: General Purpose
Wall/Panel Mount: NEMA 4X
Environment:
Indoor/Outdoor Use
Storage Temp.: -20°C to 70°C
Operating Temp.: -20°C to 70°C
Humidity:
Handheld: - 5 to 80% (non-condensing)
Wall/Panel Mount: - 5 to 80% (noncondensing)
Power:
Handheld: 9-15 VDC, 0.1 A;
Wall/Panel Mount: 95-230 VAC, ±10%,
47-63 Hz, 30 VA
Compliance:
EMC Directive: 89/336/EEC
Low Voltage Directive: 73/23/EEC
With optional HART® protocol, a
universal controller can be used in lieu
of the handheld or dedicated wall mount
IQ Links. HART option is not available
for Series 2000 Host and LinkBus
interfaces.
IQ analyzers supported by IQ Links:
CEM/O2 IQ
WDG IVCM/IQ
RM CEM O2/IQ
WDG IVM/IQ
S A/F IQ
WDG HPII/IQ
WDG Insitu/IQ
WDG HPIIC/IQ
WDG IV/IQ
WDG HPIIM/IQ
WDG IVC/IQ
WDG HPIICM/IQ
►t 20-entry event log
Thermox IQ and Series 2000-based analyzers are
EMC compliant to 89/336/EEC and Low Voltage Safety
compliant to 73/23/EEC. They are also UL and c-UL
Listed.
www.ametekpi.com
"