Model 40000 pages
THERMOX
®
PROCESS INSTRUMENTS
WDG INSITU-CENELEC
flue gas oxygen analyzer for atex ii 2 G hazardous area
Direct-sensing insitu probe suitable for applications where the flue gas temperature
does not exceed 1472°F (800°C). Completely field serviceable.
SERIES 2000 CONTROL UNIT
Display: Four-line x 20-character vacuum fluorescent.
Displays combinations of oxygen, time and date, cell temperature, user programmable text, thermocouple mV or cell
mV. Password protection, programmable pressure compensation and context-sensitive help are also provided.
Analog Output: Two isolated linear current outputs. Select
oxygen, cell temperature, thermocouple mV or cell mV. Each
output can be 4-20 mA, 0-20 mA, 20-4 mA or 20-0 mA and
is fully scalable. Hold or track during calibration and select
degree of damping. Maximum load 1200 ohms.
Alarms: Two independent oxygen alarms, each high or low
selectable. One alarm can be allocated to oxygen, calibrate or
verify. Set relays to energize or de-energize on alarm.
Contact Rating: 0.5A, 30V, 10VA max. noninductive load, AC
or DC
Diagnostics: Watchdog timer and service alarms. System test
for A/D, RAM, EEPROM and keypad. Display line 4 reserved
for full text error and diagnostic messages. Twenty entry event
log.
SENSOR
Communications: RS-485 2-way addressable
Principle of Operation: Direct-sensing insitu probe using
zirconium oxide for net oxygen measurement.
Enclosure: Approved for CE 0081,
(IP 65)
II 2 G, EEx d II C T3
Output Range: From 0-1% to 0-100% O2
Accuracy: ±1% of measured value or ±0.05% oxygen absolute, whichever is greater
Response Time:
Calibration Gas: 90% of step change in < 15 s
Process Gas: 63% of step change in < 400 s at 20 m/s (44
mph) duct velocity
Environment:
Ambient Temp.: -14°F to 122°F (-10°C to 50°C)
Relative Humidity: 0% to 80%, non-condensing
Enclosure: Approved for CE 0081,
II 2 G, EEx d II C T6
(IP65). For non-hazardous area, optional weatherproof NEMA
4 (IP56) wall/panel mount, GP (General Purpose) wall mount,
GP 19” rack mount, GP panel mount, or stainless steel weatherproof NEMA 4X (IP56) wall/panel mount. All are UL Listed
for NEC Class I, Division 2 areas.
Drift: < 0.1% of cell output per month (< 0.005% O2 per
month with 2% O2 applied)
Calibration: Oxygen cell lifetime extender. Calibrate or
verify calibration. Store last calibration and verification data.
Selectable calibration gas run time and process recovery time.
Timed automatic calibration with optional Remote Calibration
Unit.
Ambient Temp.: -20°C to 70°C (-50°F to 160°F)
Power Requirements: Nominal 115-230 VAC ±10%, 47-63
Hz; 75 VA max.
Max. Flue Gas Temp.: 800°C (1472°F)
Power Requirement: 230 VAC ±10%, 47 - 63 Hz.
(115 VAC optional); 150 VA max.
Calibration Gas: Use all calibration gases @ 0.7 kg/cm2, 1.5
to 3 L/min.
O2 Span Gas: Air or from 1.0 to 100% O2, balance N2
O2 Zero Gas: 2% or from 0.1% to 10% O2, balance N2
SYSTEM COMPLIANCE:
ATEX Directive 94/9/EC
LCIE03 ATEX 6276X (Sensor)
LCIE01 ATEX 6028 (Controller)
EN50014 and EN50018
EMC Directive 2004/108/EC
Low Voltage Directive 2006/95/EC
www.ametekpi.com
"