Apogee 5500 Series Radiation Shielded X-ray Tube with Grid Control0 pages
X-RAY
Radiation Shielded X-ray Tube
Apogee 5500 Series
Technical Datasheet
The Apogee 5500 Series is a 50kV, 50W packaged X-ray tube designed for applications
where high flux density and continuous operation are important.
Utilizing our high stability, high intensity X-ray tube technology coupled with grid-controlled variable
focus enables our Apogee design to produce very small focal spots; this makes the Apogee 5500 Series
ideal for most industrial inspection and non-destructive testing applications that require high
resolution, including PCB assembly, battery, plastic, metal and mechanical parts inspection.
Flexible and reliable, this unit is also well suited for use with X-ray optics.
The Apogee 5500 Series is configured in a compact stainless steel, leadlined package filled with dielectric oil, which enables the unit to provide
maximum X-ray shielding and heat dissipation. The design includes high
voltage and filament connectors, making it ideal for plug and play operation.
Benefits
Applications
Wide operating range enables optimal image contrast
Medical Imaging
Stable X-ray output delivers high-precision measurements
Inspection of printed circuit boards and electronic devices
Low attenuation beryllium window ensures high
Non-destructive testing of plastic, metal and mechanical
transmission of low energy X-rays
parts
Fully shielded compact package eliminates X-ray leakage
Thickness gauging
and easily integrates into your system
Analytical XRF
Specifications
Operating Voltage Range:
10-50kV
Maximum Power:
50W
Maximum Beam Current:
1.0mA
Grid Voltage:
0-100V (Oxford Shasta Power Supply recommended)
Maximum Filament Current:
1.7A
Filament Voltage:
2.0V (Nominal)
Target Material:
See product ordering table on next page
Focal Spot Size:
35µm *nominal per IEC60336, NEMA XR5-1992 (R1999)
Cone of Illumination:
22°
Spot to Window Spacing (FOD):
31.05mm ±1mm
Window Material and Thickness:
Be @ 127µm
Flux and Current Stability:
≤ 0.2% over 4-hour period
Duty Cycle:
Continuous (150 CFM airflow required)
Ambient Temperature Conditions:
Operating: 0 to 40°C
Storage: -10°C to 50°C
Humidity:
0-95% RH up to 5,000 feet
Cooling Method:
External cooling required and directed at the unit at 150 CFM. Must not exceed 55°C
at case surface.
Shielding:
0.25mR/hr @ 2" (except HV connection through HV cable)
Dimensions:
180mm L X Ø70mm (7.09" L X Ø2.76")
Weight:
1.82kg (4.0 lbs)
"