Jupiter 5000 Series Radiation Shielded X-ray Tube0 pages
X-RAY
Technical Datasheet
Radiation Shielded X-ray Tube
Jupiter 5000 Series
The Jupiter 5000 Series is a 50kV, 50W packaged X-ray tube designed for applications
where high flux density and continuous operation are important.
Utilizing our highly stable and high intensity X-ray tube technology, the Jupiter 5000 Series is ideal for medical imaging
applications and most industrial inspection and non-destructive testing applications that require high
resolution, including PCB assembly, battery, plastic, metal and mechanical parts inspection.
The 5000 Series features a stainless steel, lead-lined package that is filled with
dielectric oil, which enables the unit to provide maximum X-ray shielding and
heat dissipation. The design includes high voltage and filament connectors,
making it ideal for plug and play operation.
The Jupiter 5000 Series is available in a wide range of spot sizes, targets and
price points to meet your needs.
Benefits
Applications
Wide operating range enables optimal image contrast
Medical Imaging
Stable X-ray output delivers high-precision measurements
Printed circuit board and electronic device inspection
Low attenuation beryllium window ensures high
transmission of low energy X-rays
Non-destructive testing of plastic, metal and mechanical
parts
Fully shielded compact package eliminates X-ray leakage
and easily integrates into your system
Thickness gauging
Analytical XRF
Specifications
Operating Voltage Range:
Maximum Power:
Maximum Beam Current:
10-50kV
50W
1.0mA
Focal spot size:
P/N 93089: 50µm
P/N 93095: 55µm
Maximum Filament Current:
1.7A
Filament Voltage:
2.0V (nominal)
Focus to Object Distance (FOD):
See diagram next page
Window material and thickness:
Be @ 127µm
Cone of illumination (unobstructed):
23°
Window diameter (unobstructed):
11.43mm (.450”)
Target material:
See product ordering table next page
Target angle:
12°
Stability:
0.2% 4 hours
Polarity:
Grounded cathode
Maximum operating temperature:
55°C on case surface
Ambient operating temperature:
0°C to 40°C
Cooling method
Forced air @ 150cfm
Shielding:
0.25mR/hr @ 2” (except at HV connection)
Dimensions:
180mm L x Ø70mm (7.09” L x Ø2.76”)
Weight:
1.82kg (4lbs)
Storage Conditions:
-10°C to 55°C
Barometric Pressure: 50-106kPa; Humidity: 10-90% (no condensation)
Condensation on Be window will cause window corrosion, vacuum loss and X-ray tube failure