SD050 Drain/Air Detector Probe0 pages
DATA SHEET No. 001
Autoclave Drain
and
Air Detector/Vent Probe
Autoclave Drain and
Air Detector/Vent
Probe
Although the application conditions for the drain or air detector are not as
demanding as those presented to the autoclave chamber/load probe, their
integrity is just as critical. To maintain this integrity both types are subjected
to the same manufacturing programme as are the chamber/load probes.
S P E C I F I C A T I O N
MECHANICAL
Sheath
3.0mm OD (min) to 9.52mm OD x 100mm long (min) and up to
500mm long all in 316L (1.4404) stainless steel. When required
the probe sheath can be supplied with a welded on Tri-clamp
flange also in 316L (1.4404) stainless steel.
ELECTRICAL
Resistance Thermometer
Detector
single or duplex PT100 with 3 or 4 wire connection, Class ‘A’ or
fractional DIN tolerance and to BS EN 60751:2008.
Flying lead
PTFE insulated copper conductors, 7/0.16mm to ASTM-B-298
with a 250V RMS rating and contained within an MFA jacket with
a minimum overall length of 1000mm.
Operating range
-50 to +150°C
Thermocouple
Element
Lead wire
PTFE insulated type ‘T’ conductors, 1/0.315mm or 7/0.2mm in
Class 1 material.
Single thermocouple construction, flat pair with a PFA insulation
and jacket.
Duplex thermocouple construction, twisted pairs with a PFA
insulation and jacket.
Triplex thermocouple construction, three individual flat pairs, each
with PFA insulation and contained within a silicone rubber sleeve.
Minimum overall length of 1000mm.
Colour code
IEC colours.
Operating range
Product code : SD050
single, duplex or triplex element, type ‘T’ (Cu/Con) to BS EN
60584-1:2013.
- 50 to +150°C
For an effective and leak free entry point, sealing glands utilising soft sealant
technology are recommended. For further information please see Data Sheet No 101.
©Thermal Detection 2014