Thermocouple Entry Gland TCEG0 pages
THERMAL
Specialists in Temperature Engineering
6 Product Data Information
DATA SHEET No. 022
PRODUCT DESCRIPTION
Validation Thermocouple
Entry Gland
Validation Thermocouple Entry Gland
For
Pharmaceutical Autoclaves
Conventional methods of chamber entry for validation thermocouples allow
only limited numbers of thermocouples to be used. They are relatively time
consuming in their application and frequently prone to unacceptable
leakage.
The TCEG/1 addresses both these problem areas by providing individual
slots for 18 separate thermocouples and a rapid action, locking and
securing mechanism.
By utilising a three part sealant arrangement the gland becomes a TCEG/2
with the opportunity of increasing the capacity to 36 individual
thermocouples. Should even more thermocouples need to be
accommodated, then a 'Y' adapter connected to the chamber entry port can
double both these quantities.
Gland Body
Material
Process connection
Capacity
Thermocouple slot
S P E C I F I C A T I O N
machined from 316L (1.4435) stainless steel, round
bar stock with bore finished to 0.5 Ra (25u inch).
from 1" through to 3" Tri-clover flange or with 1" screwed
connection.
up to 36 thermocouples.
4.0mm wide.
Sealant Ring and Disc
Material
medical grade silicone and in accordance with
FDA 21 CFR 177-2600. To USP Class VI on
application.
PRODUCT CODE : TCEG
Gland Bonnet
Material
cap in CZ112 marine brass and compression nose in
CZ121 brass, both machined from round bar stock, with
spring loaded quick release locking mechanism in 316L
(1.4435) stainless steel.
© Thermal Detection Ltd. 2013
t Thermal Detection Limited Q Tel +44(0)1642 602878
Unite, O rde Wi ngate Way Fax +44(0)1642618307
Primrose Hill Industrial Estate Email tdl@thermal-detection.com