R6I-AST0 pages
R6I-AST Sensor
Integral Preamplifier
Acoustic Emission Sensor
Description and Features
Operating Specifications
PAC’s integral preamp sensors were specifically engineered
t Dynamict
t
Peak Sensitivity, Ref V/(m/s).................... 117 dB
to attain high sensitivity and have the capability to drive
Peak Sensitivity, Ref V/µbar...................... -23 dB
Operating Frequency Range...............40 - 100 kHz
Resonant Frequency, Ref V/(m/s). ............. 55 kHz
.
Resonant Frequency, Ref V/µbar................ 98 kHz
housings that are treated to minimize RFI/EMI interference.
Directionality. .................................. +/-1.5 dB
.
Care has also been taken to thermally isolate the critical
input stage of the preamplifier in order to provide excellent
t Environmental
t
Temperature Range.......................... -35 to 75ºC
temperature stability over the range of -35º to 75º C.
Shock Limit........................................... 500 g
Their integrated Auto Sensor Test (AST*) capability allows
Completely shielded crystal for maximum RFI/EMIt
immunity
long cables without the need for a separate preamplifier.
Incorporating a low-noise input, 40 dB preamplifier and a
filter all inside the sensor housing, these transducers are
completely enclosed in metal stainless steel (or aluminum)
these sensors to pulse as well as receive. This feature lets
you verify the sensor coupling and performance at any time
throughout the test.
t Physical
t
Dimensions. ................... 1.13” diameter x 1.6” ht
.
t
................................................(29 x 40 mm)
Applications
Weight. ...........................................98 grams
.
Due to its high sensitivity and low resonant frequency
Case Material. ..................... Stainless Steel (304)
.
properties, this sensor can be used for applications such
Face Material. ....................................Ceramic
.
as metal and FRP structures including pipelines or stor-
Connector. .............................................BNC
.
age tanks in petroleum, refineries, chemical plants, and
Connector Locations..................................Side
offshore platforms.
Ordering Information and Accessories
R6I. ........................................................ R6I-AST
.
Cable (specify cable length)...........................1234 - X
Magnetic Hold-Down . .................................... MHR6I
.
Amplifier......................................................AE2A
Sensors include
NIST Calibration Certificate & Warranty
Frequency response of the R6I-AST. Calibration based on ASTM
E1106; Calibration based on ASTM E976.
* AST — Auto Sensor Testing feature allows AE systems to control the sensor as a pulser
and a receiver at the same time. It can therefore characterize its own condition as well
as send out a simulated acoustic emission wave that other sensors can detect, so the
condition of the nearby sensors also can be tested.
195 Clarksville Road, Princeton Junction, NJ 08550 Phone: 609-716-4000
Fax: 609-716-0706 Email: sales@pacndt.com Internet: www.pacndt.com
A Member of MISTRAS Group Inc.
Due to continuing improvement, Physical Acoustics Corp. (PAC) reserves the right to amend specifications without notice.
Copyright © 2008 MISTRAS Group Inc. All Rights Reserved.