7059 Combined Vacuum/Pressure Calibrator0 pages
Time Electronics
7059 Pressure and Vacuum Calibrator Module
• Combined vacuum & pressure calibration
• Best accuracy 0.04% of full scale
• Pneumatic with regulator
• Calibrate gauges, transmitters, switches
• 4.5 digit display
• 5 selectable pressure units
• Over-pressure alarm
• RS-232 serial interface
• EasyCal software compatible
DESCRIPTION
The 7059 is a practical pressure and vacuum calibration and control module. It offers a controlled transition between vacuum and
pressure output and can also be used to accurately measure input pressure or vacuum. There are four available versions according
to required output range, from vacuum to 0.2, 2, 5, or 10bar.
The 7059 combines high accuracy with simple operation, and is ideal for vacuum and pressure calibration applications under 10bar.
It is an easy to use module suitable for high volume work in process plants, offshore platforms, refineries, and many other on-site
testing facilities. It is commonly used for calibration and maintenance of pressure gauges, sensors, transmitters, transducers,
indicators, switches, and more.
Features include a 4.5 digit LCD display, 5 pressure units, zeroing switch, regulated output as well as pressure measurement input.
Best accuracy is 0.04% of full scale. Separate pressure and vacuum ports are provided on the rear of the console with quick release
connections. For vacuum supply the 7149 electrical vacuum pump is available.
The module is fitted as standard with an RS-232 interface which is accessible on the rear of the CalBench console or internally
connected to a 7051 calibrator/control centre. This enables communication and read back with EasyCal calibration software (supplied
as standard with the 7051 multifunction calibrator/control centre module). With EasyCal the user can define and perform pressure
calibration procedures as well as create and print traceable calibration certificates.
SPECIFICATIONS
Range (bar)
Specifications are for 1 year at 23ºC ±5ºC. Accuracy includes linearity and long term effects.
All accuracies are ±1 digit. For operation outside specified temperature range, apply 70ppm per °C
-0.2 to 0.2
-0.95 to 2
-0.95 to 5
-0.95 to 10
Accuracy (% of fs)
0.1
0.04
0.04
0.04
Resolution (bar)
0.01mb
.0001
0.001
0.001
Units..................................................... 0.2bar: mbar, PSI, kPa, kg/cm2, cmWg / 2, 5, and 10bar: bar, PSI, kPa, kg/cm2, cmWg
Maximum Line Pressure...................... 10bar for 0.2, 2, 5bar versions / 12bar for 10bar version
Sensor.................................................. Piezo-resistive - stainless steel diaphragm
Over Pressure Warning........................ 1.2 x range full scale - audio and visual (on LCD) warning
Fittings.................................................. Quick Release
Rear Connections................................ Quick Release
Interface............................................... RS-232 for readback (using EasyCal on 7051 if ordered)
Module Width....................................... 129mm (primary or secondary console fitting)
ORDERING INFORMATION
7059..................................................... Vacuum & Pressure Calibrator Module (+ pressure range code shown below)
Range (bar)
0.2
2
5
10
Option Code
7100
7101
7102
7103
ECFLA.................................................. EasyCal Calibration Software. See separate datasheet for further information.
C178..................................................... Factory Calibration Certificate (NPL traceable)
C190..................................................... UKAS Calibration Certificate (ISO 17025)
Due to continuous development Time Electronics reserves
the right to change specifications without prior notice.
Time Electronics Ltd, Unit 11 Sovereign Way, Botany Industrial Estate, Tonbridge, Kent, TN9 1RH. United Kingdom.
T: +44 (0) 1732 355993 F: +44 (0) 1732 770312 E: mail@timeelectronics.co.uk W: www.timeelectronics.com
v1b 8/3/13
"