CS LD0 pages
FEATURES
nnnn-d^optris
nnnnHT infrared thermometers
nnnnoptnis® CS LT
nnnnSimple access into multiple sensor installations:
nnnnSmart, safe and easy for -40 to 1030°C
nnnnTemperature range: -40°C to 1030°C
nnnnResponse time: 25 ms
nnnnOptical Resolution 15:1
nnnnGreen LED alarm indication, aiming support,
nnnnselfdiagnostic or temperature code indication
nnnnUsable up to 80°C ambient temperature without
nnnncooling
nnnnSeveral outputs: 0-10 V or 0-5 V free scalable,
nnnnthermocouple type K, alarmoutput or digital output
nnnnUSB programming interface, direct serial 9.6
nnnnkBaud interface
nnnnPower supply: 5-30 V DC
nnnnGeneral Specifications | |
Environmental rating | IP 63 |
Ambient temperature | -20°C to 80°C |
Storage temperature | -20°C to 85°C |
Relative humidity | 10 - 95%, non condensing |
Vibration | IEC 68-2-6: 3 G, 11-200 Hz, any axis |
Shock | IEC 68-2-27: 50 G, 11 ms, any axis |
Weight | 58 g |
Electrical Specifications | |
Output/analog | 0-5 V or 0-10 V free scalable, thermocouple type K or alarm with adjustable voltage levels |
Output/alarm | 0-30 V/50 mA (open collector) |
3-state alarm output | adjustable threshold values and voltage level for: no alarm, pre-alarm, alarm |
Output/digital | uni/bidirectional, 9.6 kBaud, 0/3 V digital level/USB optional |
Input (0-10V) | programmable functional input for external emissivity setting/ambient temperature adjustment, triggered signal output or peak-hold function |
LED-functions | alarm indication, automatic aiming support, selfdiagnostic, temperature indication (via. temp.code) |
Cable length | 1 m (standard), 3 m, 8 m, 15 m |
Power supply | 5-30 V DC |
Current draw | 4 mA (without LEDV10 mA |
Measurement Specifications | |
Temperature range (scalable via software) | -40°C to 1030°C |
Spectral range | 8 to 14 urn |
Optical resolution (90 % energy) | 15:1 |
CF-lens (optional) | 0.8 mm @ 10 mm |
System accuracy (at ambient temp. 23 ± 5°C) | ± 1.5% or± 1.5°C1» |
Repeatability (at ambient temp. 23 ± 5°C) | ± 0.75% or± 0.75°C1> |
Temperature coefficient | ± 0.05 K/K or ± 0.05 % K3> |
NETD | 0.1 K2> |
Response time (90%) | 25 ms (adjustable up to 999 s) |
Emissivity/Gain (adjustable via 0-10 V DC input or software) | 0.100-1.100 |
Tra nsm issi vity (adjustable via software) | 0.100-1.100 |
Signal processing (parameter adjustable via software) | peak hold, valley hold, average; extended hold function with threshold and hysteresis |
Software | optris Compact Connect |
11 object temperature > 0°C; whichever is greater
nnnn2> at time constant 100 ms and 25°C
nnnnOb]
nnnn3> for ambient temperatures <18°C and >28°C; whichever is greater
nnnnInnovative Infrared Technology
"