Product Info sheet no. B1.11
M-Series
Humidity/-temperature compact sensors
with cable or plug-in connection
MCK...4S0
MCK...1K0
Technical Data
Relative Humidity
measuring range .............................................. 0...100%rh
resolution ............................................................... 0.5%rh
accuracy ......... ±2.5%rh (10...40 °C, 10...90 %rh) ±1 digit
t
other range ........... ±5% rh ±1 digit
influence of temp. ...... ±0.1% rh/K for < 10 °C and > 40 °C
Temperature
measuring ranges ........................................... colour code
resolution ............................................................. 0.125 °C
accuracy .................................. ± 0.5 K @ 23 °C ±1 digit
(details see diagramm)
General
operating voltage .............................................. 5...30 V DC
output signal :
MC(P)K1....................... 2 x 0...1 V (at RL ≥ 100 kΩ)
MC(P)K6 ................................. digital (similar I2C)
These sensors are especially adapted to the requirements of
measurement tasks which only have limited space available.
They are suitable for measuring the relative humidity and
temperature in air and other non-aggressive gases. Using
the sensors for outdoor applications is not recommended.
The sensors are based on our miniature sensors in the calHT
series (product info B1.10). They are fitted with a filter and a
4-pin plug (MCK...4S0) or with a permanently connected 1.5
m long cable (MC(P)K...1Kn). Suitable cables for the model
MCK...4S0 can be supplied in different lengths as an option.
The sensors feature high long-term stability, small hysteresis
and good dynamic performance.
series
compact sensors
M
design
with tube extension
P
direct connection plug or cable
C
phys. output
rel.humidity and temperature
K
output signal
0..1 V
1
I²C
6
none
00
seal for increased requirements
0S
measuring range H
0...100 % rh
F1
measuring range T
-30...70 °C
37
-20...80 °C
28
special edition
0...100 °C
01
-40...60 °C
-40...85 °C
46
(I²C)
48
I2C-bus-voltage ................................................. 3.3 VDC
storage temperature ........................................... -40...85°C
working temperature ............................................ -40...85°C
working temperature (with permanent cable) .... -20...70°C
power consumption ................................................... < 250 µA
degree of protection
sensor with PTFE filter ZE05.....IP40/IP64 (cable versions)
sensor with protective basket ZE07............................. IP00
sensor with protective basket ZE08 .......................... IP30
Maximum cable lengths:
tt
MC(P)K1 ................... 5 m at RL >= 100kOhm 1)
tt
MC(P)K6 ....... 0.5m at a clock rate f = 100kHz 1)
5...30 VDC
5
filter
PTFE filter ZE05
05
open ZE07 (standard)
07
with membrane ZE08
08
4-pin plug
4S0
type of connection
permanent cable 0.5 m
0K(n)
and characteristics
operating voltage
permanent cable 1.5 m
1K(n)
of the design
special lengths of cable in m
yy(n)
other specialities of design
Xyy
n: additional tube extension depending on sensor type and special design
(n) = 0:t
no tube extension
(n) = 1:t
total length sensor 53 mm
(n) = 4:t
total length sensor 95 mm
Special versions available on request.
1) t longer cables possible depending on the t
tt
ttt
load or clock rate
User instructions
Install the sensors at a place in the room where characteristic levels of humidity occur. Avoid installing them close to heaters
or windows or against outside walls.
In general, the sensors are maintenance-free. If the PTFE filter is contaminated with dust, grease and oils, this can have
a negative impact on the dynamic behaviour. In this case the sensor head of the plug version needs to be disconnected
from the cable and to be cleaned by blowing or carefully rinsing off with distilled water. After removing the filter, make sure
that the sensitive sensor surface is not touched, as this can lead to irreparable damage. An exact measured value can only
be attained again after being completely dried; this also applies to condensation.
Please consult „application instructions for the sensing elements“ (product info sheet no. A 1) for further information.
This information is based on current knowledge and is intended to provide details of our products and their possible applications. It does not, therefore, act as a guarantee of specific
properties of the products described or of their suitability for a particular application. It is our experience that the equipment may be used across a broad spectrum of applications under
the most varied conditions and loads. We cannot appraise every individual case. Purchasers and/or users are responsible for checking the equipment for suitability for any particular
application. Any existing industrial rights of protection must be observed. The perfect quality of our products is guaranteed under our General Conditions of Sale. Issue: January 2012
B1.11_E. Subject to modifications..