LR-Cal LPP 60 Pressure Test Pump 60 bar0 pages
CalibraƟon Test Pump LPP 60 (pneumaƟc)
Pressure/Vaccum ‐0,95...+60 bar (‐28inHg...+870 psi)
LPP 60
Rel. 20120525
CalibraƟon Test Pumpe LR‐Cal LPP 60, pneumaƟc operated
to generate pressure up to +60 bar (870 psi), switchable
to generate vacuum up to ‐0,95 bar (‐28 inHg)
The CalibraƟon Test Pump LR‐Cal LPP 60 is used to generate pressure
and vacuum for checking, adjusƟng and calibraƟng mechanical
and electronic pressure measuring instruments by comparaƟve
measurements. These pressure tests may be carried out in
laboratories, workshop or on site at the measuring point.
If the instrument under test and a suciently accurate
reference measuring instrument are connected to the
LR‐Cal LPP 60, the same pressure / vacuum is applied to the
two instruments when the pump is operated.
Despite ist compact dimensions, the calibraƟon test pump LR‐Cal LPP 60 is easy
to operate and allows for exact generaƟon of the required test pressures.
A change‐over switch enables the generaƟon of vacuum as well.
The LR‐Cal LPP 60 is Ʃ ed with a ne adjustment valve for the precise adjustment of pressures. The reference
instrument is screwed directly on the top of the pump and the unit under test is connected by means of a
connecƟon tube incorporaƟng an adapter 1/4" BSP rotaƟng female port with integr. sealing gasket. Sets of BSP‐,
NPT‐ and metric adapters are available as accessories. See next page.
SpecicaƟon:
Pressure generaƟon:
Vacuum generaƟon:
Pressure ports:
(standard version)
Materials:
Priming pressure:
High pressure:
Pressure adjustment:
Dimension:
Weight:
0 to 60 bar = 0 to 870 psi, switchable to
0 to ‐0,95 bar = 0 to ‐28 inHg
1/2" BSP female, rotaƟng, with integr. sealing gasket, on the top of the pump
for the reference instrument;
1/4" BSP female, rotaring, with integr. sealing gasket, at the pressure hose
for the test specimen (unit under test)
aluminium, brass, ABS, NBR
handles
volume variator
volume variator (ne adjustment control with large volume)
approx. 290 x 185 x 65 mm
approx. 960 g
Included in standard supply:
• CalibraƟon Test Pump LR‐Cal LPP 60
• Test hose, length 0.5 m
• OperaƟng manual
Recommended Reference Instruments:
± 0.025% v.E.:
LR‐Cal LPC 300 documenƟng process and pressure calibrator
± 0.025% v.E.:
LR‐Cal LPC 200 electronic pressure calibrator
±0.05% v.E.:
LR‐Cal TLDMM digital reference pressure gauge
±0.1% v.E.:
LR‐Cal TLDMM‐A01 digital test pressure gauge
±0.125% v.E.:
LR‐Cal L DM 70‐E25 digital test pressure gauge
±0.2% v.E.:
LR‐Cal TLDMM‐A02 digital test pressure gauge
±0.25% v.E.:
LR‐Cal LDM 70‐K50 digital test pressure gauge
±0.6% v.E.:
LR‐Cal LPP‐MANO‐K06 analogue reference pressure gauge
±1.0% v.E.:
LR‐Cal LPP‐MANO analogue reference pressure gauge
datasheet: „LPC 300"
datasheet: „LPC 200"
datasheet: „TLDMM“
datasheet „TLDMM‐A“
datasheet: „LDM 70"
datasheet: „TLDMM‐A“
datasheet: „LDM 70"
see next page
see next page
DRUCK & TEMPERATUR Leitenberger GmbH • Bahnhofstr. 33 • D‐72138 Kirchentellinsfurt • GERMANY
Tel. +49 (0) 7121‐90920‐0 • Fax +49 (0) 7121‐90920‐99 • DT‐Info@Leitenberger.de • www.LR‐Cal.de • www.druck‐temperatur.de