LR-Cal SOLAR0 pages
Portable Temperature Calibrator Type Solar
Dry block calibrator +200°C to +1100°C
SOLAR
Dry block Temperature Calibrator Type SOLAR & SOLAR-2I
C Temperature range: +200°C to +1100°C
C For mobile and stationary use
C Micro processor operated
C Ramp function, thermostat tests
Typical applications:
C Calibration and checking of all kinds of thermometer
C Calibration and checking of thermostats
C Computer aided calibration via RS232 interface
Technical Data:
Temperature range: +200°C to +1100°C
Accuracy of temperature reading: ±3°C
Display resolution: 0.01 / 0,1°C (switchable °C / °F)
Stability of regulated temperature: ±0.3°C at 1000°C
Mean heating time: 18°C per minute
Mean cooling time: 6°C per minute
Tubular vertical furnace with an internal quartz pipe and interchangeable equalizing block
Diameter 45 mm x dept 220 mm,
Equalizing block: Inconel 600, 44 mm diam. x 175 mm dept
Counter-current forced air cooling system, enables to check even very short probes.
Temperature uniformity: ±0.4°C horizontal and vertical
RS232 interface: supplied, build-in.
Ramp function
Block with top insulator: 4 holes (dept 155 mm), diam. 7 - 9 - 11 - 13.5 mm, others on request
Power supply: 230 VAC (optional: 115 VAC)
Power consumption: 850 VA
Case material: metall
Dimension: 450 x 170 x 330 mm
Weight: 12 kg net
Note: The stated accuracy is valid at ambient
temperature +10°C...+40°C and is guaranteed
for 1 year.
Scope of standard Supply:
C Temperature Calibrator SOLAR
C Electr. power cable
C Fuse kit
C Electr. connection cables for thermostat test
C Operating manual
C Certificate of calibration, traceable
C Block extractor tool
In addition, at type SOLAR-2I:
Two Inputs for Pt100 and thermocouples, programmable
C Pt 100 IEC 3-/4 wire, range -100/+700°C, accuracy ±0.3°C
C Thermocouple type J, range 0-1000°C, accuracy ±1°C
C Thermocouple type K/N/R/S, range 0-1300°C, accuracy ±1°C
Optional Accessories:
C Dry block in INCONEL 600, with top insulator,
holes as per customers request
C PC-Windows Software AQ2sp with RS232 cable
C Probe thermocouple typ S, optional with DKD certification
C Soft bag
The SOLAR can operate in automatic mode connected to a PC by
means of the optional AQ2sp software which enables to carry out
probe calibrations and cyclical life tests; test results can be stored
and printed, so they are easily traceable in compliance with
ISO 9000 standards.
DRUCK & TEMPERATUR Leitenberger GmbH
Bahnhofstr. 33 C D-72138 Kirchentellinsfurt C Germany
Tel.: +49 - 71 21 - 9 09 20 - 0 C Fax: +49 - 71 21 - 9 09 20 - 99
E-Mail: DT-Export@Leitenberger.de C http://www.druck-temperatur.de
Technische Änderungen vorbehalten. Freibleibend C (Rel. 100121) • All technical modifications reserved. Without engagement