Differential pressure transmitter in cleanroom-conform panel design - testo 63830 pages
We measure it.
nnnntesto 6383
nnnnDifferential pressure transmitter in cleanroom-conform panel design
nnnnSPECIFICATIONS
nnnnSPECIFICATIONS
nnnntesto 6383
nnnnMeasurement of differential pressure; optional:
nnnnhumidity and temperature
nnnnAutomatic zero-point adjustment guarantees high,
nnnntemperature-independent accuracy and long-term
nnnnstability
nnnnLow measurement range up to 10 Pa ensures
nnnnhighest precision at lowest pressures
nnnnFlat housing allows flush surface integration in the
nnnncleanroom wall
nnnnDisplay with multi-language operating menu and
nnnnoptical alarm display
nnnnEthernet, relay and analog outputs allow optimum
nnnnintegration into individual automation systems
nnnnSelf-monitoring of the transmitter and early warning
nnnnfunction guarantee high system availably
nnnnThe P2A software for parameterization, adjustment
nnnnand analysis saves time and costs in
nnnncommissioning and maintenance
nnnnScalability of ±50 percent of the measuring range
nnnnfinal value and free scalability within the measuring
nnnnrange
nnnntesto 6383
nnnnThe differential pressure transmitter testo 6383 was
nnnndeveloped specially for monitoring low differential
nnnnpressures in the measuring range from 10 Pa to 10
nnnnhPa. In cleanroom technology, the maintenance of
nnnnpositive pressure prevents the entry of contaminated
nnnnair in critical zones. Thanks to an optional internal or
nnnnexternal probe from the probe series 6610, the
nnnnadditional recording of humidity and temperature
nnnnwith one instrument is also possible.
nnnnThe testo 6383 is particularly outstanding thanks to
nnnnthe automatioc zero-point adjustment which ensures
nnnnhigh accuracy and long-term stability.
nnnnThe integrated self-monitoring and early warning
nnnnfunction also guarantees the operator high system
nnnnavailability.
nnnnAreas of application:
nnnn• Monitoring positive and negative pressure in
nnnncleanrooms, operating theatres and isolation
nnnnrooms
nnnn• Optional monitoring of humidity and temperature
nnnnin cleanrooms
nnnnConfigurable alarm management with adjustable
nnnnresponse delay and alarm acknowledgement
"