QSG 750-II0 pages
QSG 750-11
Optical Emission Spectrometer
for Ultimate Performance
Metal Analysis
, 3000 -
2500 -
Al Intensity die
I Technical Specification*
1. Optical System
- Paschen-Runge line-up
- temperature stabilised
2. Vacuum system
- automatic vacuum control
- pump duty cycle < 5%
3. Spark stand
- optimized for low Ar consumption
- patented self cleaning
- spark frequency up to 1 kHz
4. Spark generator
- Gated Digital Source (GDS)
- completley maintenance free
- fully semiconductor-based & digital control
5. GISS-Functions
- time-resolved spectroscopy
- online processing of single spark intensities
- elimination of line interferiences
- spectral background reduction
and optimized limits of detections
- selectable integration windows
for each installed channel
- determination of soluble/insoluble components
- free definable algorithms for the
determination of inclusions
6. Software Functions
- Windows™ Software
- automatic precision control & averaging
- automatic reprofiling
- type calibration
- charge control
- data module for statistic process control
7. Applications
- all matrices
8. Installation
- dimensions approx. 119 x 130 x 90 cm (wxhxd)
- weight approx. 550 kg
- permissible operating temp. +10 to +40°C
- argon supply: 3 bar, Ar 4.8 or better
- power connection 230V, 50/60Hz, 1.5 kVA
3
Our QSG 750-11 spark emission!
spectrometer can also be called
OBLF's flagship model. This
single- or multi-matrix model is ideal for
all applications demanding the lowest
possible detection limits, the highest
degree of reproducibility and additional
metallurgical information about the
analyte. The application spectrum goes
from primary producers like steel mills
or other smelteries through metal:
processing companies right up to
research institutes and universities.
At first glance, the QSG 750-H is
comparable with the
QSN 750-11 model. Given
identical dimensions, it is also fitted
with tried-and-tested hardware
components like the temperature-
stabilised 750 mm vacuum optics, the
Gated Digital Source (GDS) technology
with freely definable parameters and the
patented, self-cleaning spark stand. An
exhaust filter ensures consumed argon is
purged. The easy accessibility of all key
device parts simplifies service tasks that
need to be performed by the user.
The main difference to the other
OBLF spectrometers lies in the
data logging system, which is
based on the GISS technology (Gated
Integration of Single Sparks) developed
by QBLF. In combination with logging
and online processing of the single spark
intensities of all measuring channels,
time-resolved spectroscopy gives rise to
many additional applications. Single
spark spectrometry (or PDA pulse
distributed analysis), for instance,
provides critical information about the
microscopic composition of the sample.
In this respect, key words are the
detection of inhomogeneous samples,
differentiation between dissolved and
undissolved components (e.g.
aluminium in steel) as well as the
detection and determination of
inclusions. To ensure best possible use
can be made of these options, the
OBLFwin software not only includes all
standard functions (materials control,
automatic program selection, etc.), but
also provides additional features that
permit simple definition of parameters
when a bad sample is detected or when
analysing inclusions. All the single spark
data can be made available for export to
other computer systems. Time-resolved
spectroscopy can reduce spectral
interference and substantially improve
the detection limits of many elements in
comparison to standard spectrometers.
This opens up completely new avenues
when analysing pure metals.
All of the QSG 750-11 model's
many functions are available for
manual and automated
operation, as many applications have
shown in the steel industry in particular.
In such cases, sample handling is
performed by a robot and the counter-
electrode is cleaned by a special unit
fitted to the spark stand. The measured
values are automatically transferred to
superordinate systems using a network
connection. Integrated software modules
control and monitor the functionality of
the spectrometer. These functions also
include analysing control and
recalibration samples.
Contact: OBLF Ges. fOr Elektronik & Feinwerktechnik mbH, Salinger Feld 44, D-58454 Witten,
Tel. +49 2302 9879-0, Fax: +49 2302 9879-80, Email: info@oblf.de, Internet: www.oblf.eu
"