QSN 750-II0 pages
QSN 750-11
Optical Emission Spectrometer
for Universal, High Precision
Metal Analysis
Our QSN 750-n model is a
spark emission spectrometer
that can be used as a single-
or multi-matrix system thanks to its
750 mm vacuum optics. This results
in a broad field of application, which
not only covers the entire range of
metal producing and processing
enterprises, but also includes a large
spectrum of options for material
testers and testing institutes. Both the
integration system and the
photomultipliers, which are arranged
in accordance with customer
specifications, are housed inside the
optics and are therefore protected
against external influences. The
temperature-stabilised optics and the
Gated Digital Source (GDS)
technology guarantee a high degree
of signal reproducibility and thus
analytical constancy. The easy-to-
reach self-cleaning and patented
spark stand and aperture window
enable problem-free maintenance by
the user. An exhaust filter ensures the
consumed argon is purged.
In contrast to the compact
GS 1000-11, lie QSN 750-11 can
also be used within automated
production environments. In such
cases, the spectrometer is combined
with a sample-preparation device and
a sample-handling robot to form a
single unit, which naturally also
ensures automatic cleaning of the
counter-electrode. The analysis
results are transferred online to the
superordinate computer system using
a network connection. In order to
guarantee optimum quality control,
the spectrometer initiates any
necessary control sample readings at
certain time intervals, whereby the
robot is again responsible for sample
handling. With the help of previously
defined limiting values, the
spectrometer automatically monitors
control samples. Key system
functions are permanently monitored
and if necessary this information is
transmitted to other systems.
Alarge number of analysis
programs can arise especially
when using multi-matrix
systems to perform extensive
analytical tasks. For this reason, the
matching spectrometer software,
OBLFwin, provides numerous
individual programming options to
perfectly suit user requirements. In
addition, the automatic program
selection option simplifies
assignment of the right analysis
program. Depending on the
application, the measured values
appear on screen approx. 16-30
seconds after the start of the
measuring process and can then be
printed out, transferred within the
network or evaluated later.
Technical Specifications
1. Optical system.
- Paschen-Runge line-up
- temperature stabilised
2. Vacuum system
- automatic vacuum control
- pump duty cycle < 5%
3. Spark stand
- optimized for low Ar consumption
- patented self cleaning
- spark frequency up to 1 kHz
4. Spark generator
- Gated Digital Source (GDS)
- completley maintenance free
- fully semiconductor-based & digital control
5. Software Functions
- Windows09 Software
- automatic precision control
- automatic reprofiling
- automatic averaging
- type calibration
- charge control
- data module for statistic process control
6. Applications
- all matrices
7, Installation
- dimensions approx. 119 x 130 x 90 cm (wxhxd)
- weight approx. 550 kg
- permissible operating temp. +10 to +40°C
- argon supply: 3 bar, Ar 4.8 or better
- power connection 230V, 50/60HZ, 1.5 kVA
Contact:
OBLF
Ges. fur Elektronik & Feinwerktechnik mbH
Salinger Feld 44 D-58454 Witten
Tel. +49 2302 9879-0 Fax: +49 2302 9879-80
Email: info@oblf.de Internet: www.oblf.eu
"