Tailin Glove Integrity Tester HTY-GIT010 pages
Glove Integrity Tester
HTY-GIT01
Glove Integrity Tester
HTY Glove Integrity Tester is
designed to check the integrity of Isolators
/ RABS systems gloves / sleeve or long
gloves. It is artistic and efficient. HTY
glove integrity tester includes a control
panel that compatible with clean room and
easy to operate, a control system, as well
as a sealed cover matched with the
isolator operating hatch. It can work in
positive pressure mode. Test results will
be displayed in numbers to the operator.
Based on the reference values provided by
the manufacturer, operator can ascertain
the integrity of detecting gloves.
♦ Features
It complies with the ISO 14644-7
standards. It has a stainless steel welded
base, and equipped with four casters, two
can twist, two fixed. The tilt control panel
help you easy to observe and operate. It
also has an ergonomic mobile handle.
Printing capabilities to deliver results is
convenient to save records.The SIEMENS
touch screen enable you to control the
machine easily. Leak testing can be
conducted before or after running the
isolator.
HTY-GITO1 Glove Integrity Tester is a dedicated
apparatus which is used to check the integrity of
Isolators / RABS systems gloves / gloves or one-piece
gloves.
HTY-GITO 1 glove integrity tester includes an
control panel that is easy to use, a control system, as
well as a test port matched with the isolator operating
hatch.
HTY-GITO 1 glove integrity tester can operate in
positive pressure mode. Test results will be shown as
pressure drop values within unit time. Based on the
reference values provided by the manufacturer,
operator can ascertain the integrity of detecting
gloves.
♦ Features
Portable design, easy to use;
2. Wholly stainless steel chassis, easy to
clean, suitable for use in clean areas;
3. Tilt control panel, easy to observe and
operate;
4. Test gas is external compressed air and
nitrogen;
5. With results printing , easy to save
records;
6. Adopts SIEMENS programmable controller
(PLC) and touch screen to control, endowed with a
friendly man-machine communication interface;
7. Leak testing can be conducted before or
after running the isolator.
8. Complies with ISO 14644-7 standards.
♦ technical parameters
Input pressure: 0.2MPa - 0.6MPa
Inflation Pressure: 500 - lOOOPa
Test time: 5 ~ 8min
Pressure display resolution: 0.1 Pa
Detection accuracy (smallest aperture): 500 ji m
Instrument Weight: 20kg
Host Dimension: (H*W*H) 530mm*235mm
*330mm
Test Port Specifications: 0255mm (other
specifications can be customized)
Printing: dot matrix mini printer
Data storage: 10 sets
Power supply: 220V, 50Hz
Power: 200W
"