Continuous Gas Analyzers, extractive
OXYMAT 6
1
General information
Special versions
■ Overview
Special applications
Besides the standard combinations, special applications concerning the material in the gas path and the material in the sample chambers are also available on request.
TÜV version QAL
As a reference variable for emission measurements according to
TA-Luft, 13th and 17th BlmSchV
■ Design
The function of the OXYMAT 6 gas analyzers is based on the
paramagnetic alternating pressure method and are used to
measure oxygen in gases.
■ Benefits
• Paramagnetic alternating pressure principle
- Small measuring ranges (0 to 0.5 % or 99.5 to 100 % O2)
- Absolute linearity
• Detector element has no contact with the sample gas
- Can be used under harsh conditions
- Long service life
• Physically suppressed zero through suitable selection of
reference gas (air or O2), e.g. 98 to 100 % O2 for purity monitoring/air separation
• Open interface architecture (RS 485, RS 232, PROFIBUS)
• SIPROM GA network for maintenance and service information
(option)
• Electronics and physics: gas-tight isolation, purging is possible, IP65, long service life even in harsh environments (field
device only)
• Heated versions (option), use also in presence of gases condensing at low temperature (field device only)
• EEx(p) for zones 1 and 2 according to ATEX 2G and ATEX 3G
(field device only)
■ Application
Fields of application
• For boiler control in incineration plants
• For safety-relevant applications (SIL)
• In the automotive industry (testbed systems)
• In chemical plants
• For ultra-pure gas quality monitoring
• Environmental protection
• Quality monitoring
• Versions for analyzing flammable and non-flammable gases
or vapors for use in hazardous areas
1/128
Siemens AP 01 · 2015
19" rack unit
• With 4 HU for installation
- in hinged frame
- in cabinets with or without telescopic rails
• Front plate can be swung down for servicing purposes
(laptop connection)
• Internal gas paths: hose made of FKM (Viton) or pipe made of
titanium or stainless steel (mat. no. 1.4571)
• Gas connections for sample gas inlet and outlet and for reference gas: fittings, pipe diameter of 6 mm or ¼"
• Flow indicator for sample gas on front plate (option)
• Pressure switch in sample gas path for flow monitoring
(option)
Field device
• Two-door enclosure with gas-tight separation of analyzer and
electronics sections
• Each half of the enclosure can be purged separately
• Analyzer unit and piping can be heated up to 130 °C (option)
• Gas path and stubs made of stainless steel (mat. no. 1.4571)
or titanium, Hastelloy C22
• Purging gas connections: pipe diameter 10 mm or 3/8"
• Gas connections for sample gas inlet and outlet and for reference gas: clamping ring connection for a pipe diameter of
6 mm or ¼"
Display and control panel
• Large LCD panel for simultaneous display of:
- Measured value (digital and analog displays)
- Status bar
- Measuring ranges
• Contrast of LCD panel adjustable using menu
• Permanent LED backlighting
• Washable membrane keyboard with five softkeys
• Menu-driven operation for parameterization, test functions,
adjustment
• User help in plain text
• Graphic display of concentration trend; programmable time
intervals
• Bilingual operating software German/English,
English/Spanish, French/English, Spanish/English,
Italian/English