WP400 - Spectrum analysis 1 Hz - 400 kHz0 pages
WP400 Probe
1 Hz - 400 kHz
Electric & Magnetic field measurament
Isotropic & True RMS measurement
Spectrum analysis probe
Measurements in accordance with
International Standards
1 Hz - 400 KHz
E-H
Power grid
Spot measurement
of E and H at
transformer stations
and high-voltage lines
Railway
Measurement of E and H
fields generated in trains
or near railway facilities
Industry
Measurement in manufacturing facilities with strong
electromagnetic fields to
ensure worker´s safety
RMS
ISOTROPIC
Technical specifications
1 Hz - 400 kHz
Isotropic, RMS
Combined measurement of electric and magnetic field
Field strength mode (results in V/m, µT, etc.)
WPE (Weighted Peak Exposure) Evaluation
Results displayed in % of the selected standard
FFT-Based Real-Time Spectrum Analysis
Results in V/m, µT, etc. and % of the selected standard
Exposure limits
(probe in weighted mode)
Public and occupational ICNIRP 2010, 2013/35/EU Directive
Customizable to other standards
Measurement range
Iso
170
25
dited c
cre
Measurement mode
025 ·Ac
17
· ISO
lab
Sensor type
ration
alib
Frequency range
Field Strength Mode
E field: 1 V/m - 100 kV/m
H field: 50 nT - 10 mT
WPE Evaluation
E field: 0.5 % - 200 % (Occupational)
H field: 0.5 % - 200 % (Occupational)
FFT function (Selective analysis)
E field: 4 mV/m - 100 kV/m (Occupational)
H field: 0.5 nT - 10 mT (Occupational)
Linearity
± 1% (typ.)
± 2% (max.)
Isotropic response
± 5% (typ.)
Calibration
ISO 17025 accredited (ILAC)
Calibration period
24 months (recommended)
Operating temperature
-15ºC to 35ºC
Dimensions
270 mm x 115 mm Ø
Field sensor area
100 cm2
Weight
210 g
Safety, Quality, Service
+34 93 320 80 55
wavecontrol.com
info@wavecontrol.com