Mini-spectrometer0 pages
MODULE
Mini-spectrometer TM series
C10082MD, C10083MD
For UV to near IR, integrating optical system, image sensor and circuit
TM series mini-spectrometers are polychromators integrated with optical elements, an image sensor and a driver circuit. Two models are
available: C10082MD (TM-UV/VIS-MOS) and C10083MD (TM-VIS/NIR-MOS). Light to be measured is guided into the entrance port of TM series
through an optical fiber and the spectrum measured with the built-in image sensor is output from the USB port to a PC for data acquisition. No
external power supply is required since USB bus power is used for circuit operation. The TM series comes with sample software, device driver
and DLL. The sample software lets you easily set measurement conditions, acquire and save data, and display graphs.
Features
Applications
l High throughput due to transmission grating made of quartz l Light source spectrum measurement
l Sunlight or illumination analysis
l Highly accurate optical characteristics
l No external power supply required: Uses USB bus power l Absorbance measurement
l Wide spectral response range
l Easy to install into equipment
l Wavelength conversion factor *1 is recorded in internal memory
l Supports external trigger input *2
s Optical characteristics
Parameter
Spectral response range
Spectral resolution Max.
(Spectral response half width) * 3
Wavelength reproducibility * 4
Wavelength temperature dependence
Spectral stray light * 3, * 5
s Electrical characteristics
Parameter
A/D conversion
Integration time
Interface
USB bus power current consumption
TM-UV/VIS-MOS
C10082MD
200 to 800
TM-VIS/NIR-MOS
C10083MD
320 to 1000
6
8
nm
-33
nm
nm/°C
dB
±0.2
0.02
-35
Value
16
5 to 10000
USB1.1
100
Unit
nm
Unit
bit
ms
mA
s General ratings / Absolute maximum ratings
Parameter
Value
Unit
Dimensions
94 (W ) × 90 (D) × 55 (H)
mm
Image sensor
CMOS linear image sensor (S8378-1024Q)
Number of pixels
1024
pixels
Slit * 6
70 (H) × 800 (V)
µm
Optical NA
0.22
Connector for optical fiber
SMA905D
Operating temperature * 7
+5 to +40
°C
Storage temperature
-20 to +70
°C
*1: A conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. A calculation
factor for converting the A/D converted count into the input light intensity is not provided.
*2: Coaxial cable for external trigger input sold separately. Refer to the “Mini-spectrom eters Selection Guide” for details on
external triggers.
*3: Depends on the slit opening. Values were measured with the slit opening listed in the table.
*4: Measured under constant light input conditions
*5: W hen monochromatic light of the following wavelengths is input, spectral stray light is defined as the ratio of the count
measured at the input wavelength, to the count measured in a region of the input wavelength ±40 nm.
C10082MD: 500 nm, C10083MD: 650 nm
*6: Entrance slit aperture size
*7: No condensation
1