RVO/U-L4 Flow monitor and indicator for gases0 pages
Flow Monitor Flow Indicator
RVO/U-L4
|t^^^l/min|
<v»n<16°
100
<-20
I bar 1
l^^m zoo*
350^
Range of Application
Operating Data
Operation
• Float measuring principle
Application
• Mechanical engineering
• Medicine technology
• Pharma industry
• Chemical industry
• Research and development
• Cooling systems and cooling circuits
Features
• Universal orientation
• High reliability
• High switch accuracy
• Infinitely variable switchpoint adjustment through user
• Scales are burned into the sight glass
• Threaded connection, special threads on request
Installation hints
• The operating instructions for RVO/U-L must be observed!
• Download: www.meister-flow.com
Operating pressure max.
Pressure drop
Maximum temperature
Accuracy
Measuring Ranges
16 bar
0,02-0,2 bar
100 °C (optional 160 °C)
±10 % of full scale
Type
RVO/U-L40001
RVO/U-L40002
RVO/U-L40003
RVO/U-L40005
RVO/U-L40008
RVO/U-L40012
RVO/U-L40014
RVO/U-L40020
RVO/U-L40024
RVO/U-L40035
RVO/U-L40042
Switch range for Air at 1 barabs. and 20 °C <1)
[Nl/min]
0,2 - 1,3
[SCFH]
0,4 - 2,75
[SCFM]
0,5
0,8
1,5
2
3
3,5
5,5
7
10
10
2
3
5
8
12
14
20
24
35
42
1,05
1,70
3,5
4,5
6,5
7,5
12,0
15,0
21,0
21,0
4,25
6,40
10,5
17,0
25,0
29,5
42,0
50,0
74,0
89,0
(1' The specified data are switch-off points,
other switch ranges on request.
-meisters
strdmungstechnik §>
"