SP2300-H0 pages
Embracing Challenge
Gas Sample Probe Series SP®
Version SP2000, SP2000-H, SP2300-H, SP2400-H
SP2000-H
Special Features
Optimum operational reliability
Universal applicability
Adaption to nearly all process condi-
tions due to its compact and modular
design
Easy mounting and maintanancet
Small volume, fast response time
Application
The M&C sample probes version SP2000...,
SP2300-H and SP2400-H are used for continuous xtraction of gases from dust loaden,
e
hight temperature and/or humid processes.
Description
Depending on the gas composition, it is possible that the standard material of the probe
body (stainless steel 316) will not be sufficiently corrosion resistant. In this case probe
SP2300-H made of PTFE or SP2400-H made of
Titanium should be used.
The sample probes are designed for easy
installation, reliable operation and trouble
free maintenance. They are versatile in application and depending on the task to be
performed various sample tubes or pre-filters
(see data sheet 2.14 and 2.17), not included in
the scale of delivery, can be simply screwed
into the probe (G 3/4» thread).
The filter element (ceramic is standard,
others optional) with a large surface area
and high capacity is located in the external
housing.
The design offers little or no stagnant space external to the process. The
probes are designed so that changing
a filter element does not involve the use
of tools. In this operation neither the sample probe tube nor sample line need to
be removed, thus avoiding contamination
of the clean gas path and maintaining the
i
ntegrity of the system.
The sample tube can be cleaned, washed or
rodded-out from outside the process. The
special design of the heating element of the
SP2000/2300/2400-H (with protective hood)
p
ermits controlled heating of the complete
filter housing, including the mounting flange
up tp 180 °C (Version /H320 up to 320 °C).
This ensures reliable operation xternal to the
e
process preventing the temperature alling
f
below the dew-point.
The temperature of the standard probe is
controlled by an integral compact design
capillary sensor hermostat with an excess
t
temperature limit switch and an alarm function for emperature failure. The probe can be
t
provided with test and reference gas facilities.
Technical specifications and illustrations are without
obligation, subject to modifications. 10.04/06.06
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.1
"