Series MBU153 - 150W Medical Power Supply0 pages
Series MBU153
150W PFC MEDICAL POWER SUPPLY
FEATURES
l Open frame power supply
150W
l output
Single
l operating voltage 90-260VAC, 47-63Hz
Wide
Active
l Power Factor Correction (PFC)
ANSI/AAMI ES 60601-1:2005
l
(UL/cUL 3rd Edition) EN 60601-1:2006
(TUV/T-mark 3rd edition)
Input
l to output : 2MOPP
(Means of Patient protection)
Class
l I
SPECIFICATION
Operating Voltage Range: 90 ~ 260VAC
Input Frequency: 47 ~ 63Hz
Input Current (Low Line): at full load, Vin = 100VAC
1.74~1.76A
Input Current (High Line): at full load, Vin = 240VAC
0.71~0.73A
Low Line Inrush Current: at full load, 25°C, Cool start Vin =
115VAC, 50A (max.)
High Line Inrush Current: at full load, 25°C, Cool start Vin =
230VAC, Max. 100A
Power Factor Correction: at full load, Vin = 240VAC; 0.95 ~ 1
Over Voltage Protection: 112 ~ 132%
Over Current Protection: 110 ~ 150%
Load Regulation: Vin = 230VAC; 3 ~ 5%
Line Regulation: at full load, 0.5 ~ 1%
Time of Transcient Response: Io = full load to half load
Vin = 100VAC, Max. 4ms
Hold-Up Time: at full load, Vin = 110VAC, Min. 20ms
Start-Up Time: at full load, Vin = 100VAC, Min. 2s
Ripple & Noise (Peak to Peak):
Full load, Vin = 90VAC, Max. 1%
Safety Approvals Input Voltage Range: 100 ~ 240VAC
Safety Ground Leakage Current:
Vin = 240VAC / 60Hz, Max. 0.25mA
Temperature Coefficient: All output, +/- 0.04%/°C
Operating Altitude: Up to 3000m
Dielectric Withstanding Voltage:
Primary to secondary; Min. 6990VDC
Dielectric Withstanding Voltage for Primary to PE:
Primary to PE, Min. 2121VDC
Isolation Resistance: Test voltage = 500VDC, Min. 50M Ohm
Efficiency: Io = full load, Vin = 230VAC, 84 ~ 89%
Storage Temperature: -40 ~ 85°C
Operating Humidity: 0 ~ 95%
Storage Humidity: 0 ~ 95%
Operating Temperature at 25°C; Calculated per MIL-HDBK-217F:
Min. 100.000 Hrs
Warranty: 24 months