Process Instruments
nnٕ Large-volume furnace extends maintenance intervalspreventing plugging and failing Simplified sample delivery system
Օ Passive cooling system avoids complex heat management
Utilizes proven, patented high temperature reactor systemՕ Industrial design withstands the most severe conditions Analyzer-protecting advanced diagnostics
nnReduced Cost of Ownership and Maintenance
nnThe patented large-volume furnace prevents severe-duty samplesfrom plugging or failing prematurely consequently extending routinemaintenance intervals. A platinum catalyst provides an enlarged
surface area for the oxidation reaction in the furnace. This increases
the lifecycle of the catalyst reducing cost of ownership. nn
Simplified Sample Delivery System
nnInnovative sample delivery system avoids inherent mechanicalbreakdown of complicated sample injection mechanisms.
Customarily high temperature analyzers use an injection valve
mechanism,which is prone to fail due to its narrow passages, small
parts and exposed seals.The astroTOC HT has a continuous sample feed by a peristalticpump providing a robust, easy-to-maintain sample injection into the
furnace. This dramatically reduces maintenance cost. nn
How To Order
nnContact your local Hach sales representative to configure a TOC analyzer for your application. Primary ApplicationsՕ Industrial Water Wastewater >
nnSpecifications*Range Sample Temperature Range Drain Connection Note: Must specify range at time of order. 2պ to 70C (36ں to 158F) 1 1/2-inch OD standard drain pipe 0-25 up to 0-20,000 mg/L TOC Operating Temperature Range Carrier Gas Accuracy 5ں to 40C (41ں to 104F) Clean, CO >
2 free air or Nitrogen at 2.8-6.2 bar (40-90 psig) ڱ5% of reading at ranges less than 1000 mg/L with and without dilution at 25C (77ںF)2% of reading in the range of 2000 to 20,000 mg/L with dilution at 25ѺC (77F) Recorder Outputs Two 4-20 mA analog outputs selectable for sampleconcentration, analyzer system warning or auto
range indication Compliance/Certification Alarms CE certified, listed to UL and CSA safety standards by ETLStandard Methods 5310 B, EPA 415.1 Repeatability ڱ5% of reading at ranges less than 1000 mg/L with and without dilution at 25C (77ںF)2% of reading in the range of 2000 to 20,000 mg/L with dilution at 25ѺC (77F) Five alarms selectable for sample concentrationalarm, analyzer system warning or analyzer system
shutdown alarm. Enclosure Cold Rolled Steel epoxy powder coated, IP54/NEMA 12Optional Stainless Steel IP54/NEMA 12 Approximately 983 mm (38.7 inches) tall, 973 mm(38.3 inches) wide, 244 mm (9.6 inches) deep Optional Serial Communication Minimum Detection Limit Dimensions One multi-function RS232 or RS485 optional serialport (MODBUS < 0.1 mg/L for the range 0-25 mg/L at 25ںC (77F) >
ڮ , CSV) Mounting Response Time Wall mount Power T90 ≤ 8 minutes (includes TIC sparging) 115/230 Vac 50/60 Hz (switch selectable),1500 VA maximum Shipping Weight Inlet Pressure 212 lbs. (97 kg) 0.15-6 bar (2-87 psig) Sample Inlet/Outlet Connection Flow Rate 1/4-inch OD tube, compression fitting >
*Subject to change without notice. 20-200 mL per minute >