A6500-TP - Temperature / Process Card0 pages
Specifications Sheet
CSI 6500 Machinery Health™ Monitor
January 2016
A6500-TP - Temperature / Process Card
The A6500-TP Temperature Process Card is a component of the CSI 6500-ATG Machine
Protection System. The card is equipped with 4 measuring channels for temperature
measurements with thermocouples or resistance temperature detectors (RTDs) and
for measurements of process signals 0 - 1 V, 0 - 10 V or 0/4 - 20 mA standard signals.
Besides this, the card contains 2 digital inputs and 6 digital outputs. The measured
signals are transmitted through the internal RS 485 bus to the A6500-CC Com Card
and converted to Modbus RTU and Modbus TCP/IP protocols for further transmission
to host computers or analysis systems. In addition, the Com Card provides the
communication through the USB socket at the face plate for the connection to PC/
laptop for the configuration of protection cards and for visualization of measuring
results. Apart from that, the measuring results can be output through analog outputs
0/4 - 20 mA. These outputs have a common ground and are electrically isolated from
system supply. The operation of the A6500-TP Temperature Process Card is performed
in the A6500-SR System Rack, which also provides connection of supply voltages and
signals. The A6500-TP Temperature Process Card provides the following functions:
QQ Temperature measurement with RTDs
QQ Temperature measurement with thermocouples with internal and external cold
junction compensation
QQ Measurement of standard process signals 0 - 1 V, 0 - 10 V or 0/4 - 20 mA
QQ Acquisition of digital signals with two digital inputs
QQ Output of analog values for each measuring channel through 4 current
outputs 0/4 – 20 mA
QQ Output of alarms through 6 digital outputs, the assignment to the measuring
channels can be configured in any way
Signal Inputs
Number of Inputs
Four (Up to Eight in Temperature Mode) independently configurable for different temperature
sensors or process signal inputs, nonreactive, short
circuit proof.
Input Resolution
24 bit ADC
Isolation
The four inputs are galvanically isolated against each
other based on levels of IEC 60204-1,
respectively IEC 61131-2
Voltage Process Inputs
Range
0 to 10 V
Accuracy
±1% of full scale
Impedance
>100 kΩ
Temperature Drift
±0.5% of full scale within operating temperature range
of -20°C to +70°C
www.emersonprocess.com/mhm
QQ Two-channel, 3U size, 1-slot plugin module decreases cabinet space
requirements in half from traditional
four-channel 6U size cards.
QQ API 670 compliant,
hot-swappable module.
QQ Self-checking facilities include
monitoring hardware, power input,
hardware temperature, sensor,
and cable.
QQ Use with 2,3 or wire RTD´s,
thermocouples, or analog process
input s