HD32.7-HD32.8.8-HD32.8.16 DATALOGGER0 pages
HD 32.7
HD 32.8.8
HD 32.8.16
TECHNICAL SPECIFICATIONS
Number of Inputs
8 DIN 45326 8-poles male connectors.
0.01°C (in the range ±199.99°C)
0.1°C in the remaining range
Measuring range
-200°C … 650°C
Display
Backlit graphic LCD 128x64 pixel.
Keyboard
15 keys, configurable also without PC.
Keyboard locking function
with password.
Memory
divided into 64 blocks.
Memory capacity
96.000 storages for each one of the inputs.
Security of data stored
Temperatura
°C, °F, °K
Resolution
24
1min/month max drift
Unit of measurement
HD32.7
8 INPUTS DATA LOGGER FOR pt100 SENSOR PROBES
The instrument HD32.7 is a robust 8 inputs data logger for Pt100 sensor temperature
probes equipped with SICRAM module. 4 wires Pt100 Probe.
•tUnit of measurement °C, °F, °K configurable.
•tFlash memory organized in 64 sections with a total capacity of 96.000 acquisitions
for each one of the 8 inputs. Storage can be managed in two ways:
-t when the available memory is full, data are overwritten by starting from the
oldest ones (circular memory),
-t storage stops when the available memory is full.
•tSimultaneous display of the 8 inputs.
•tMaximum, minimum or average of the stored values.
•tSelectable storage interval: 2, 5, 10, 15, 30 seconds, 1, 2, 5, 10, 15, 20, 30
minutes and 1 hour.
•tData logging: instantaneous or postponed, with the possibility of selecting the
storage start and end.
•tData download: RS232C, 1200...38400 baud or USB 1.1 – 2.0.
•tDeltaLog9 software for data download and processing.
•tLCD backlit graphic display 128x64 pixel.
•tInstrument setup through the keyboard; no connection required to the PC.
•tSecurity password for keyboard locking.
•tPower supply: 4 1.5V alkaline C-BABY type batteries or external power supply
12VDC-1A.
•tConsumption @6VDC:
<60μA t when the instrument is off
<60μA t in sleep mode with 8 probes connected
<40mA t during data logging with 8 probes connected
•tUse of the HD32.7 data logger: in the field for machine or equipment measurements,
plant or machine testing, production check, oven mapping.
±0.01°C ±1digit (in the range ±199.99°C)
±0.1°C ±1digit in the remaining range
Internal watch accuracy
HD32.7 - HD32.8.8 - HD32.8.16
DATALOGGER
Instrument accuracy
when storingt
unlimited
Power supply
4 per 1.5V alkaline Batteries type C-BABY
External 12Vdc-1A power supply.
Connector, external ∅ 5.5mm, internal ∅ 2.1mm.
Current consumption @6Vdc
Autonomy
<60μA when the instrument is off
<60μA in sleep mode with 8 probes connected
<40mA during data logging with 8 probes connected
200 hours with 7800mAh alkaline batteries and 8
probes connected
Data download
12 Vdc 1A
RS232C from 1200 to 38400 baud, galvanically
isolated. Sub D 9-pole male connector.
USB 1.1 – 2.0 galvanically isolated.
USB 1.1 - 2.0
RS232C