Horsman-joint RTD temperature probes TM1204(1)0 pages
3GGON
Stock Code:430519
Horsman-joint RTD temperature probes (TM1204)(1)
►Applications:
Heat exchanger units, Refrigeration and testing equip-
ments.
^Technical data:
Protection tube: Stainless steel 304/316
Sensor element: R100, R500, R1000 meets IEC 751
Tolerance: Class B: At=±(0.30 + 0.005 |t| )'C(standard)
Class A: At=±(0.15 + 0.002 |t| )'C
Transmitter: Integrated temperature transmitter 24VDC, 4~20mA, 2-wire
Transmitter errors: Standard ±0.5%. ±0.2% is optional.
Response time: \>a approx. 10secs, in water 0.2 m/sec, diameter 6mm
►Wiring diagram:
1x2-wire 1x 3-wire 1x4-wire
white red white red red white white red red
Z
RTD
2x4-wire
white
red
white red
RTD
-j~TT
RTD
RTD
51
BOGON
Stock Code:430519
►Dimensions:
►Order Details:
TM1204 | (1) | (2) | (3) | (4) | (5) | (6) | / | (7) |
(1 protection tube diameter D
4 4mm
5 5mm
6 6mm
8 8mm
9 9mm
11 11mm
(2)lnsertion Length EL
50 50mm
100 100mm
250 250mm
500 500mm
Note: other lengths as per customers' requirements.
(3)Process connection G
000 No
123 M16x1.5
Details in "Thread code table"
(4) Sensor element
11 1 x Pt100 in 2-wire circuit
51 1 x Pt500 in 2-wire circuit
61 1 x Pt1000 in 2-wire circuit
13 1 x Pt100 in 3-wire circuit
14 1 x Pt100 in 4-wire circuit
12 2 x Pt100 in 4-wire circuit
16 2 x Pt100 in 6-wire circuit
62 2xPt1000 in 4-wire circuit
(5)Measuring range/Connecting cable
13 -50'C ~+ 180 'C/Silicone
20 -40'C~+400'C
Specification: the actual operating range is from -20'C
to +120'C with integrated transmitter.
(6) Cable length L (L>100mm)
100 100mm
200 200mm
2000 2000mm
Note: other lengths as per customers' requirements.
(100mm<L<1000mm, L=50mm, L>1000mm,
L.=100mm)
(7)Additional options
00 No
02 Shielded connecting cable (D>4mm)
06 Output 2-40mA in 2-wire
07 Ferrules as per DIN 46 228 part4 (Standard)
08 Multipole connector (Providing types)
99 Special options
List the additional options in sequence, separated by
commas.
52
"