Ion Implanted Silicon Alpha Particle Detectors0 pages
fc^K BAL
nnnnF BALTIC SCIENTIFIC
nnnn21 INSTRUMENTS
nnnnSilicon Ion Implanted
nnnnAlpha Particle Detectors
nnnnApplication
nnnnIon Implanted Silicon Alpha Particle Detectors is a product for the precise alpha spectroscopy. The
nnnnthin entrance window of the detector provides good energy resolution even in close location of the
nnnnalpha radioactive source and also provides high efficiency registration of alpha particles.
nnnnFeatures
nnnn• The detectors can operate without hermetization due to location of P-N junction inside of the detector
nnnncrystal
nnnn• Contacts are formed using ion-implantation method and provide thin, well-formed junction
nnnn• Relatively thin dead layer (less than 500 A)
nnnn• High solidity entrance window
nnnn• Possibility working in vacuum
nnnn• Possibility of annealing the detectors up to 100 °C
nnnn• The detectors may be equipped with BNC or MICRODOT connectors adapted for different customer
nnnnneeds.
nnnn• The detectors are manufactured with open window as well as with metalized window.
nnnnBaltic Scientific Instruments
nnnnGanibu Dambis 26
nnnnRiga, LV-1005
nnnnLatvia
nnnnPhone: (+371)67383947
nnnnFax: (+371)67382620
nnnnEmail: detectors@bsi.lv
nnnnwww.bsi.lv
nnnn40 0 & P particle
nnnnBringing Solutions
nnnnSpecification
nnnnDetector
nnnnarea, mm2
nnnn450
nnnn300 -1200**
nnnnGuaranteed energy
nnnnresolution*, (keV)
nnnnAlpha Beta
nnnn20
nnnn30
nnnn15
nnnn17
nnnnDetector
nnnntype
nnnnOpen
nnnnMetalized
nnnnOpen
nnnnMetalized
nnnnThickness, pm
nnnn400 +/- 30
nnnn400 +/- 30
nnnn400 +/- 30
nnnn400 +/- 30
nnnnDetector bias
nnnnvoltage, V
nnnn50
nnnn50
nnnn* Shaping time - 1 us.
nnnn**The detectors of other sizes are available.
nnnnThe Ion Implanted Silicon Alpha Particle Detectors with metalized coating of entrance window can be
nnnnused as a part of radioactive aerosol monitors. This is special version of the detectors having following
nnnnfeatures:
nnnn• Allows the detector operation in ambient light
nnnn• The metal coating provides mechanical and chemical protection. The thickness of entrance window is
nnnnless than 2 [jm
nnnn• Opportunity of operation at bias voltage - from +15 to +24 V.
nnnn■fLSpe. ti^uneADA 11 ■ [>n. i«iFnt*^iKl^hi^t'.SXZMM vl)eU<tQp\Trip_U+50V_RI_iZZ+gen*hn- < ZS (I*-2Q12 *5Z:*S r>< Alpha >]
nnnnFfc Analyser Options Ptoc«slng Teals YWnttows HsJp
nnnnH||■> g rff ||A-0,-^1 Hi, JL,C||» • = m\\l A
nnnn^■1* "I
nnnn238Pu, 239Pu, 233U
nnnnspectrum
nnnn182 LNfl! 2O7.0O jfls* 20?.DC |Ds*tl 0.00% Ouwud: 1S») \Caris.
nnnnI E/lftffly; 8230.19 FW; H81
nnnnAlpha Spectrometers include:
nnnn• Vacuum chamber with vacuum control system,
nnnn• Preamplifier,
nnnn• Amplifier,
nnnn• Discriminator,
nnnn• High voltage supply,
nnnn• Test-generator
nnnn• Digital multichannel analyzer.
nnnnAlpha spectrometer same as vacuum control are operated by software which also allows to process
nnnnacquired spectra and analyze measurement results.
nnnna &p particle 41
"