HDT/Vicat Allround_cool0 pages
Product Information
HDT/Vicat Allround with integrated cooling unit
HDT/Vicat 6300 Allround, motorized
test weight application
Range of applications
Determination of Heat Deflection Temperature (HDT) on
thermoplastics, hard rubber, as well as fiber reinforced
and filled hardenable plastics, according to ISO 75
Parts 1 to 3 and ASTM D 648 Method A and B
Determination of Vicat Softening Temperature (VST)
on thermoplastics, according to ISO 306 and ASTM D
1525
Determination of creep test characteristics
under deflection loading
PI 603 2.0312
Description
The HDT/Vicat Allround instruments have been designed for comfortable testing in research and development, incoming goods control, product monitoring as
well as for education and training purposes. Various
instruments with up to 6 testing stations are available.
testXpert® II enables convenient parametrization of the
test sequence and display of test values.
Automatic test sequence
As soon as the immersion liquid has reached the
starting temperature, the specimens are placed in the
testing stations and manually lowered into the bath.
The next step is to place the test weight, called for by
the standard, manually or optionally motorized, and
then start the test sequence from the PC Software.
After pretempering while under load, the travel signals
are set to zero by the program, and the heating of the
immersion liquid is started with the preset heating rate.
As soon as the required flexure, edge fiber extension
or penetration is achieved at all stations, the heating is
switched off and the test is ended.
The cooling of the immersion liquid is started automatically by the program via magnetic valve and then the
specimens can be taken out of the bath.
Functions, elements and interfaces
• t Builtin SPScontrol for temperature regulation and
test data acquisition
•t Cooling controlled by magnetic valve
•t Motorized immersion of the testing stations
•t Manual or optional motorized placing of the test
weights
•t Safety thermostat
•t Electronic bath level monitoring
•t Test control and data acquisition via PC with
testXpert® II (RS232Interface required in PC)
•t Integrated compensation of the thermal expansion
Cooling of the immersion liquid
The instruments are equipped with an integrated cooling unit for short cooling times.
Specimen data
VST specimen dimensions, max.: 10 x 6.5 x 10 mm
HDT specimen dimensions, max.: 13 x 15 x 130 mm
HDT support distances: 64 and 100 mm
(optional 101.6 mm)