XR AirJet Sample Coolers0 pages
Performance Specifications
XR402
XR902
-40 °C
-90 °C
Standard
Optional
Temperature Control Range
-40 to 100 °C
-90 to 100 °C
Temperature Control Stability
± 0.5 °C
± 0.5 °C
0.1 °C
0.1 °C
Flow Rate Range
0.25 to 2 scfm
(7 to 56 lpm)
0.25 to 2 scfm
(7 to 56 lpm)
Communication Interface
(Control Only)
RS485/USB
(with software)
RS485/USB
(with software)
Maximum Low Temperature
Temperature Control
Temperature Indication Display*
Note: Performance specifications are based on SP Scientific test data from units using
dry air with dew point of less than -100° and operating at an ambient room temperature
of approximately 22 °C (72 °F).
XR902 with Probe
Temperature at Flow †
XR402
XR902
XR902
(Control)
Features
(No Control)
Mechanically refrigerated.
0.25 scfm (7 lpm)
-30 to 100 °C
-65 to 100 °C
-75 °C
Temperature control range of -90 to 100 °C.
0.50 scfm (14 lpm)
-33 to 100 °C
-80 to 100 °C
-85 °C
Flow rates to 2 scfm (56 lpm).
8’, 15’, 20’ or 35’ delivery line lengths available.
0.75 scfm (21 lpm)
-37 to 100 °C
-85 to 100 °C
-90 °C
Low flow safety buzzer.
1.00 scfm (28 lpm)
-40 to 100 °C
-90 to 100 °C
-90 °C
Includes USB adapter, 2m cable and software
for plug and play operation.
1.25 scfm (35 lpm)
-40 to 100 °C
-90 to 100 °C
-90 °C
Multi-function temperature controller.
1.50 scfm (42 lpm)
-40 to 100 °C
-90 to 100 °C
-90 °C
1.75 scfm (49 lpm)
-40 to 100 °C
-90 to 75 °C
-90 °C
2.00 scfm (56 lpm)
-40 to 75 °C
-90 to 50 °C
-90 °C
Benefits
Eliminates the need for expendable refrigerants.
Satisfies a wide range of applications.
Satisfies specific application requirements.
Non-magnetic delivery line.
Alerts users when air flow to the unit is low.
Permits complete control of the machine from
a PC and allows for data logging.
Provides accurate temperature control with
temperature indication.
Air Requirements
XR402
Flow
Minimum Pressure
Dew Point
XR902
56 lpm (2 scfm)
56 lpm (2 scfm)
45 psi
45 psi
< -60 °C
< -100 °C
Electrical Requirements
XR402
XR902
XR902
(Control)
Agency Approvals
(No Control)
60 Hz Option
120 V, 5 A‡
120 V, 12 A‡
120 V, 11 A‡
50 Hz Option
CE (2006/42/EC), (2006/95/EC), (006/108/EC)
220 V, 2 A‡
220 V, 5 A‡
220 V, 4 A‡
Note: All equipment configured with the 60 Hz option is supplied with a 5-15P NEMA
configuration (i.e.,
). All equipment configured with the 50 Hz option is supplied with
a non-terminated AWG 18-3 power cord.
3538 Main Street, Stone Ridge, NY 12484 USA 845.255.5000
800.431.8232 Fax: 845.687.7481 www.SPScientific.com/ftsthermal