PO2-150 pages
Product Specification Sheet
PO2-15 Percent Oxygen Sensor
1.25
.93
Sensing Surface
.76
Specifications:
Sensor Technology t
Signal Output1t
Measuring Ranget
Micro Fuel Cell
78 - 122 uA
0 - 100 Percent
Response Time T90 t
Accuracy Full Scale2t
Repeatability F.S. t
Temp Coefficientt
Operating Temp t
Storage Temp t
Recommended Flow t
Humidityt
13 Seconds
± 1% of Signal
± 0.5%
2.54 % / º C
0 to 50 º C
0 to 45 º C
0.5 - 2 SCFH
0 - 100% RH t
(Non - Condensing)
34 - 38 Months
6 Months
12 Months
CenterFoilNegative t
Outer Foil Positive
Expected Life3t
Storaget
Warranty4t
PCB Connectiont
PCB Assembly
1.00 .53
.53
Direct Replacement Guide
Analytical Industries: PSR-11-25
Teledyne: B-5
1. Signal output is measured in air at 25 º C, sea level.
2. Accuracy full scale is calculated with constant pressure and temperature & proper calibration (90% O2
value on full scale range). Drastic temperature change
can result in a maximum error of ± 10%.
3. Expected life is calculated when O2 is 20.9% @ 25 º C,
sea level.
4. Southland Sensing warrants these sensors for the
period noted above to be free from defects in materials
and workmanship. Southland Sensing will not be held
liable for sensors damaged due to customer neglect.
4045 E. Guasti Rd. #203
Ontario, CA 91761 USA
Web: www.sso2.com
Phone: 1-949-398-2879
Fax: 1-949-315-3622
E-mail: sales@sso2.com