MIC-10k10 pages
Index: WMGBMIC10K1 (MIC-10k1)
WMGBMIC5050 (MIC-5050)
INSULATION RESISTANCE METER
MIC-10k1 / MIC-5050
WIRELESS DATA TRANSMISION
Insulation resistance measurement:
· up to 40 TΩ for MIC-10k,
· up to 20 TΩ for MIC-5050,
- measurement voltage any in the range of:
· 50...5000 V for MIC-5050 (50...1000 V at 10 V and 1...5 kV at 25 V),
· 50…10000 V for MIC-10k1 (50...1000 V at 25 V and 1...10 kV at 50 V),
- continuous indication of measured insulation resistance or leakage current,
- automatic discharge of measured object capacitance voltage after the end of insulation resistance measurement,
- acoustic signaling of 5 seconds intervals to facilitate capturing time characteristics,
- adjustable measuring time to 99'59'’,
- metered T1, T2 and T3 test times for measuring one or two absorption coefficients from the range of 1…600 s,
- polarization index (PI), absorption coefficients Ab1, Ab2 and dielectric absorption ratio (DAR) measurement,
- indication of actual test voltage during measurement,
- 1.2 mA, 3 mA and 5 mA test current,
- insulation resistance measurement using two- or three-wire method,
- measurements with test leads up to 20 m...
- protection against measuring live objects,
- automatic measurement of multiple core cables with the additional AutoISO-5000 adapter (for MIC-10k1 max. voltage 5 kV)
- measurement of capacitance during the measurement of RISO.
- measurement of temperature (with additional probe - WASONT1),
- step voltage insulation resistance measurement (SV),
- Dielectric Discharge calculation (DD),
- location of damage (burnout),
Sonel S.A.
Wokulskiego 11
58-100 Świdnica, PL
tel. +48 74 85 83 860
fax +48 74 85 83 809
export@sonel.pl
www.sonel.pl
- digital filters function for measurements in high noise enviroment
- continuity measurement of protective connections and equipotential bonding in accordance with EN 61557-4 with current > 200 mA,
- adjustable limits for measured resistance RISO and RCONT,
- measurement of leakage current during insulation resistance testing,
- DC and AC voltage measurement in the range of 0...750 V,
- drawing graphs on the display during measurement,
- innovative memory with possibility of description of: measurement points, facilities, names of customers.
- operating with mini Bluetooth keyboard (option),
- graphic LCD 5,6'’ backlit,
- keyboard backlit,
- power supply from main power line or battery packs,
- built-in fast charger,
- the instruments meet the requirements of the EN 61557 standard.