Isco 710 Ultrasonic Flow Module0 pages
Isco 710 Ultrasonic
Flow Module
Uses non-contacting sensor
Plug in a 710 Flow Module and turn your
6700 Series or Avalanche® Sampler into
a combination sampler and flow meter.
The Ultrasonic sensor is mounted above the
flow stream, and it transmits a sound pulse
that is reflected by the surface of the flow
stream. The elapsed time between pulse and
echo determines the level of liquid in the
channel. The liquid level is then converted
into flow rate by the sampler controller.
A built-in temperature probe in the sensor
automatically compensates for changes in air
temperature to ensure measurement accuracy.
Because its sensor does not contact the liquid,
the 710 is easy to install and gives you longterm
dependability with no scheduled
maintenance. The 710 is often the best choice
for flow streams with corrosive chemicals or
high concentrations of grease, suspended
solids, or silt.
Standard Features
.. Non-contacting ultrasonic sensor
is unaffected by flow stream
composition
.. Built-in flow conversions for most
applications, including weirs and
flumes, Manning formula,
data points, or equations for
special situations
.. During the program’s operation,
current flow and level values
are viewable on the sampler’s
LCD display
.. All level data stored in the sampler
is available for later retrieval,
reporting, and graphing using Isco
Flowlink® software
Applications
.. Flow and level measurement in
streams containing harsh chemicals,
grease, or suspended solids
.. 6712 or Avalanche sampler triggering
based on level or flow
.. Flow-proportioned sample collection
.. Treatment capacity analysis
.. Stormwater monitoring
.. Combined sewer overflow studies
.. Long-term river and stream gauging Simply plug in one of the environmentally-sealed
modules to expand monitoring capabilities. They