MODEL 3000PB Process Percent Oxygen Analyzer0 pages
TELEDYNE
ANALYTICAL INSTRUMENTS
Ambient Air Calibration
The high accuracy and fast response of Models 3000PB and
3010PB is only surpassed by its ability to calibrate without the
assistance of support gases. The Micro-fuel Cell produces an
output that is linear from 0% to 100% O2 enabling the use of
ambient air for calibration. These units can also accommodate
the use of span or zero gas when necessary. The 3000PB and
3010PB offer stainless steel sample tubing and a nylon cell holder.
Advantages
MODEL 3000PB
Process Percent
Oxygen Analyzer
T
eledyne’s Model 3000PB and 3010PB Bulkhead
Mount Percent Oxygen Analyzers are versatile,
microprocessor-based instruments for detecting
percent oxygen in a variety of gases.
The 3000PB serves as a general purpose unit.
The 3010PB is designed as a “split architecture”
instrument, meaning that a general purpose, wall
mounted Control Unit for non-hazardous areas
controls a remote Analysis Unit that can operate in
hazardous areas.
Simple menu choices, membrane command switches
and a large LED display make setup and operation of
these instruments simple and fast.
Three Analysis Ranges
Three user-configurable ranges are standard,
with excellent linearity precluding recalibration
when changing ranges. Two fully programmable
concentration alarms and relay functions provide
the versatility to satisfy nearly any requirement.
All features offer a sophistication that assure these
instruments will provide years of service.
Convenient Outputs for Data
Standard 0-1 VDC outputs are provided for oxygen
measurements and range identification. An
RS-232C serial port interface allows the user to
introduce span and zero gases through a host
computer.
•tLinearity of analysis across three user-selectable ranges
•tRemote access to all functions from computer
•tExtended-life, maintenance free sensor
•tAmbient air calibration
Standard Features
•tNEMA 4 enclosure
•tRFI hardened
•tSecurity code protection
• Versatile analysis over a wide range of applications
•tMicroprocessor based electronics
•tProgrammable Auto Ranging
•tTwo fully-adjustable concentration alarms and a system failure
alarm
•tAir calibration range for convenient spanning at 20.9%
•tSelf diagnostics
Options
-Ct Electrically
operated
cal/zero
valves
-Vt Vacuum
service
-St Stainless
steel cell
holder
-Mt Isolated 420 mADC signal output
and range ID
Model 3010PB
Split Architecture
-Ft Flame arrestors for Group C&D service t(3010PB only)
-Ht Flame arrestors for Group B service (3010PB only)
Built for Reliability and Performance