Analyzers for Oxygen Compounds T4000 pages
^Bfi^TELEDYNE
ADVANCED POLLUTION INSTRUMENTATION
Model T400 UV Absorption 03 Analyzer
The Model T400 UV Absorption analyzer uses a system based
on the Beer-Lambert law for measuring low ranges of ozone in
ambient air.
A 254 nm UV light signal is passed through the sample cell
where it is absorbed in proportion to the amount of ozone
present. Periodically a switching valve alternates measurement
between the sample stream and a sample that has been
scrubbed of ozone. The result is a true, stable ozone
measurement.
All T Series instruments offer an advanced color display
capacitive touch screen, intuitive user interface, flexible I/O, and
built-in data acquisition capability. All instrument set up, control
and access to stored data and diagnostic information is
available through the front panel, or via RS232, Ethernet, or USB
com ports either locally or by remote connection using the
included APIcom™ software.
►► Ranges: 0-100 ppb to 0-10 ppm,
user selectable
►► Single pass ultraviolet absorption
Dual ranges and auto ranging
►► Large, vivid, and durable color
graphics display with touch screen
interface
►► Ethernet, RS-232, and (optional)
USB com ports
►► Front panel USB connections for
peripheral devices and firmware
upgrades
►► 8 analog inputs (optional)
►► Adaptive signal filtering optimizes
response time
►► Temperature & pressure
compensation
►► Internal programmable zero, span
and precision points (optional)
►► Comprehensive internal data
logging with programmable
averaging periods
►► Ability to log virtually any operating
parameter
►► Two-year warranty
Free Customer Support by telephone and email for the life of the instrument
"