T7010 pages
TELEDYNE
ADVANCED POLLUTION INSTRUMENTATION
A Teledyne Technologies Company
T SERIES
JgSSo POliimON INSTRUMEHIM10N
Model T701 Zero Air System
The T701 adds Modbus TCP/IP connectivity to the proven
reliability and performance of our zero air systems. This allows
remote monitoring of key instrument functions including
dewpoint status, scrubber temperature, box temperature, and
output pressure. Additionally, this digital access provides the
ability to remotely disable the internal pump, allowing longer
service intervals and less power consumption. The user can also
remotely start instrument warm-up prior to arrival at the site.
The T701 is a fully self-contained source of clean, dry air for
dilution calibrators. It can also be used to create purge air for
permeation tube ovens or burner air for FID analyzers. The T701
includes a dewpoint sensor, an oil and diaphragm-free pump, and
scrubbers to remove SO2, NO, NO2, O3, and H2S. Optional high
performance scrubbers are available to remove CO and
Hydrocarbons.
►► Modbus Connectivity
►► Dewpoint Sensor
►► Once/year maintenance interval
►► Regenerative, heatless dryer for
maintenance-free water removal
►► Long-life scrubbers for SO2, NO,
NO2, O3 and H2S
►► Optional CO or CO/Hydrocarbon
scrubber
►► Automatic water drain
►► Automatic pump control with demand
tank
►► Two-year warranty
The regenerative, heatless dryer removes water and produces
output with a dewpoint of less than -20°C (up to 15 SLPM flow
rate). The system's pressure, temperature, and dew point values
are all continuously monitored, which allows a microcontroller to
adjust the pump cycling frequency, valve timing, and heater
power for optimal performance.
Free Custom&r Support by t&l&phon& and email for the life of the instrument
"