GasShield® Gasketfilter Instruction Sheet0 pages
slot. Bend the tab back to secure the tag to the line. (Illustration ‘B’ - Metal
Tag)
10. Helium leak test connections for integrity.
ISO 9001:2008 CERTIFIED
11. Close vent line valve if installed.
12. Open upstream and downstream valves.
Mott GasShield® GasketFilter
FLOW
INSTALLATION AND REPLACEMENT INSTRUCTIONS:
Warning: It is the installer’s responsibility to comply with appropriate safety
procedures for working with gases under pressure and for toxic and other hazardous
gases. If installer is unsure of proper procedures, appropriate authority must be
consulted prior to beginning installation and/or removal of a filter.
Caution: GasShield GasketFilters are preconditioned with clean, dry nitrogen.
Extreme care should be taken to minimize the time which elapses between the
removal of the packaging and the installation of the filter to minimize contamination
of the filter.
Caution: Like any VCR compatible gasket, the GasShield GasketFilter is intended for
one-time use. Re-use of the GasketFilter could result in system leaks.
1.
Close valves upstream and downstream of the filter.
2.
Perform cycle purge to the gas system to get the component to be removed,
inerted and depressurized.
3.
Disassemble the connection and remove the used filter or gasket.
4.
Inspect all sealing surfaces and clean or replace as needed.
5.
In order to avoid contamination of the filter media, do not handle the
Illustration ‘A’ - Suggested Installation
Illustration ‘B’ - Metal Tag
SPECIFICATIONS:
Retention:
Greater than 99.9999999% (9 LRV) removal of
all particles measured down to 0.003 micron.
Operating Conditions:
Catalog No.
Rated Flow SLPM
GSG-V4-1-N
GSG-V4-3-N
GSG-V4-1-2
GSG-V4-1-S
GSG-V4-3-S
GSG-V4-1-2S
1
3
20
1
3
20
Material
Nickel
Nickel
Nickel
Stainless
Stainless
Stainless
Maximum Differential Pressure at 20°C
Forward Flow:
1000 psid
Reverse Flow:
1000 psid
Maximum Operating Temperature:
450°C Inert Gases
ready for immediate installation.
Gasket Surface Finish:
10 RA Microinch
Remove the filter from the bag. Make certain that the filter will be installed so
Recommended Changeout Interval:
System Dependent
GasketFilter without wearing clean gloves. Do not open the inner package until
6.
that the flow direction agrees with illustration ‘A’.
Note:
7.
The GasketFilter is effective in either flow direction.
Immediately insert the filter while being careful not to damage the connection
sealing surfaces. DO NOT INSTALL ADDITIONAL GASKETS. Tighten the
connection.
8.
Perform cycle purge to the gas system to get the installed component inerted
and free from atmospheric contamination.
9.
Install the metal tag provided with each filter on the gas line, adjacent to the
connection, by forming the tag around the line, and inserting the tab into the
GSKTFILT.DOC
REV 7 – 01/10
0.4 Micron Filtration
Steel
Steel
Steel 0.4 Micron Filtration