WDG IV0 pages
WDG-IV flue gas oxygen analyzer
THERMOX
®
PROCESS INSTRUMENTS
Close-coupled extractive design for fast response in a wide range of flue gas
applications up to 3000°F (1648°C). Completely field serviceable.
SENSOR
Principle of Operation: Zirconium oxide
for net oxygen measurement.
Output Range: From 0-1% to 0-100%
Accuracy: ± 0.75% of measured value or
± 0.05% oxygen, whichever is greater
Response: 90% of a step change < 6 sec.
Drift: < 0.1% of cell output per month
(< 0.005% O2 per month with 2% O2
applied)
Aspirator Air Requirements: 10 to 20
scfh (4.7 to 9.4 L/min) at 15 to 100 psig
(1 to 7 kg/cm2)
Max. Flue Gas Temp / Probe Type:
1300°F (704°C) / 316 SS
1875°F (1024°C) / 310 SS
3000°F (1648°C) / Hexoloy®
Probe Lengths:
SS: 24” - 108” (60 cm - 271 cm)
Hexoloy®: 24” - 72” (60 cm - 182 cm)
Max. Sample Dewpoint: 392°F (200°C)
standard. High dewpoint option for sample
dewpoints up to 700°F (371°C).
Sample Pressure: ± 10 in. water gauge
Environment:
Ambient Temp.: -5°F to 160°F (-20°C to
71°C); -5°F to 140°F (-20°C to 60°C) with
Div. 2 Option
Relative Humidity: 10% to 90%, noncondensing
Enclosure: Lift-off NEMA 3R, weather
resistant, stainless steel. Optional hinged
NEMA 4X (IP56), explosion-proof, purged,
and floor mount versions available.
Power Requirements: 115 VAC, ± 10%,
47-63 Hz, 600 VA max. (650 VA max. floor
mount option); 230 VAC, ± 10%, 47-63 Hz,
1850 VA max. (1900 VA max. floor mount
option)
Calibration Gas Requirements: Use
calibration gases @ 10 psig, 1.5 scfh
(0.70 kg/cm2, 0.7 L/min.)
O2 Span Gas: Air or from 1.0 to 100% O2,
balance N2
O2 Zero Gas: 2% or from 0.1 to 10% O2,
balance N2
SERIES 2000 CONTROL UNIT
Display: Four-line x 20-character vacuum
fluorescent. Displays combinations of
oxygen, time and date, cell temperature,
user programmable text, thermocouple
mV or cell mV. Password protection,
programmable pressure compensation and
context-sensitive help are also provided.
Analog Output: Two isolated linear current
outputs. Select O2, cell temperature,
thermocouple mV or cell mV. Each output
can be 4-20 mA, 0-20 mA, 20-4 mA or
20-0 mA and is fully scalable. Hold or track
during calibration and select degree of
damping. Maximum load 1200 ohms.
Alarms: Two independent oxygen alarms,
each high or low selectable. One alarm
can be assigned as oxygen, calibrate
or verify. Set relays to energize or deenergize on alarm.
Contact Rating: 0.5 A, 30V, 10 VA max.
noninductive load, AC or DC
Diagnostics: Watchdog timer and service
alarms. System test for A/D, RAM,
EEPROM, and keypad. Display line 4
reserved for full text error and diagnostic
messages. Twenty-entry event log.
Communications: RS-485 2-way
addressable
Environment:
Ambient Temp: 14°F to 122°F (-10°C to
50°C)
Relative Humidity: 10% to 80%, noncondensing
Enclosure: Standard weatherproof NEMA
4 (IP 56) wall/panel mount. Optional GP
(General Purpose) wall mount, GP 19” rack
mount, GP panel mount, or stainless steel
weatherproof NEMA 4X (IP 56) wall/panel
mount. All are UL Listed for NEC Class I,
Division 2 areas. Purged and explosionproof versions also available.
Calibration: Oxygen cell lifetime extender.
Calibrate or verify calibration. Store last
calibration and verification data. Selectable
calibration gas run time and process
recovery time. Timed automatic calibration
with optional Remote Calibration Unit.
Power Requirements: Nominal 115-230
VAC ± 10%, 47-63 Hz, 75 VA max.
System Compliance:
EMC Directive 2004/108/EC
Low Voltage Directive 73/23/EEC
EXCESS FUEL OPTION
Extends operating range of analyzer
from excess oxygen only to include
substoichiometric conditions (excess
fuel). Allows two-point calibration in
excess fuel range. Measure, display, and
provide alarms and analog outputs as
follows:
t
Display Options: Excess fuel,
combustibles, oxides/ fuel, fuel/
oxides, combined excess oxygen/
excess fuel (combustibles)
t
Display Range: 0-50% excess fuel
t
Output Range: 0-1% to 0-50% excess
fuel
t
Alarms: Standard alarms can be used
for high or low excess fuel levels
www.ametekpi.com
"