High Performance Digital Force Gauge ZT series0 pages
IMADA CO., LTD..
Standard Model Digital Force gauge ZTS series
Standard Model Digital Force Gauge ZTS series
FORCE MEASUREMENT
Feature 1 Clear display / Easy setting.
Feature 2 You can capture peak value precisely due to high sampling.
Feature 3 Easy data management on PC with included software, ZT-Logger.
Feature 4 Various functions to improve usability in different environment.
Data Display Gauge Setup Information
Monitor Data
• Start | |
Capture Value | ■ Stop |
Statistics
Maximum
Minimum
No.60
-3.44
1.6596
Average
Data Qty. |78
Std. Dev. 13.2445
Number of +NG [F
1
Number Force | Go/ | Time | Date | |
49 1 331N | 17:50:10 | 2013/11/07 | ||
50 6.37N | OK | 17:50:10 | 2013/11/07 | |
51 | 4.38N | 17:50:11 | 2013/11/07 | |
52 | 8.60N | OK | 17:50:11 | 2013/11/07 |
53 | 4.45N | 17.50:11 | 2013/11/07 | |
54 | 5.68 N | OK | 17:50:11 | 2013/11/07 |
55 | 5.87N | OK | 17:50:11 | 201311/07 |
56 | 5.76N | OK | 17:50:11 | 20131107 |
57 | 2.84N | NO | 17:50:11 | 2013/11/07 2013 11 07 |
58 | 7.32N | OK | 17:50:11 | |
59 | 3.40N | 17.50:11 | 2013 11 07 | |
60 | 11.03N | + NG | 17.50.11 | 20131107 |
61 | 1.66N | 17:50:11 | 2013 11 07 | |
62 | 0.84N | 17:50:12 | 2013/11/07 | |
63 | j | 17:50:12 | 2013 11 07 |
i£) Copy to Clipboard
5
How to use
ED) | 0/100 | |
0.0 | 1 | |
Hi + 200 0 Lo- | 200.0 | |
i | FOACE MEASUREMENT iutnt |
Included Software ZT-Logger
Feature 1 Clear display / Easy setting
- Clear organic display has excellent
readability and reduces misreading.
- Setting menu with multiple languages helps
easy operation.
Program Menu
High Low Setpoints
Peak Functions ^
Internal Memory
* Higher range of 2500N is different in shape
Simply handheld use
High repeatability
with a test stand
Make graphs to analyze
force transition
Feature 2 You can capture peak value precisely at high sampling.
Press peak button
once, so that it
holds peak value.
It reduces missing load value due to high sampling speed (2000 data / sec). Peak value can be accurately captured.
Sampling speed: 2000Hz
->High repeatability
Breaking tests of the same sample
Sampling speed: 100Hz
■>Dispersion of the measured data
1)1 i | 1 i |
Pata measured at 2000Hz
Ave. of peak : 33.2N
Nearly the same value
[Data measured at 100Hz
Ave. of peak: 29.92N
Values vary. Ave. is lower.
Depending on sample or testing conditions, you may not get the same result.
With optional software, Force Recorder, you can make graphs of force versus time.
For further information, go to "Related products" on page 5
1/8
2015/07/28
"