Sterisensor0 pages
Sterisensor® Autoclave Chamber/Load Sensor
Sterisensor® withstands the environmental
conditions found in pharmaceutical autoclave
chambers. Sterisensor’s engineered moisture
seal prevents the ingress of moisture which is
the leading cause of failure.
Thermocouple Type T or RTD, 100Ω@0ºC,
.00385Ω/Ω/ºC calibrations are available,
sensor lengths and diameters are configured
to meet application requirements.
Mechanical
Sensor Diameter:
•t 3.0mm, 3.18mm (.125”), 4.76mm (.187”),
6.0mm, 6.35 (.250”)
Sensor Length:
•t 75mm (3”) minimum, 305mm (12”) maximum
Sheath Material:
•t 316 Stainless Steel
Leadwire/Jacket Length:
•t 900mm (36”) length minimum
RTD
Element:
•t Single or dual PT 100 with 3 or 4 wire
connection, Class ‘B’, and ‘A’ tolerance meeting
BS EN 60751:1996. Fractional 1/10 tolerance
available.
Leadwire/Jacket:
•t PTFE insulated 26AWG stranded conductors
with thick walled extruded silicone rubber jacket
Operating Range:
•t -50ºC to +150VC (-58ºF to +302ºF)
•t Vacuum to 3.5 BAR (50 PSIG)
Note: Multiple detector configurations are
determined by the sensor diameter.
Thermocouple
Type:
•t Type T: Single or Dual to ASTM E230
Leadwire/Jacket:
•t PTFE insulated 24AWG stranded conductors
with thick walled extruded silicone rubber jacket
Operating Range:
Sterisensor® with chisel point
•t -50ºC to +150ºC (-58ºF to +302ºF)
•t Vacuum to 3.5 BAR (50 PSIG)
Accessories
•t Various types of leadwire sealing gland
assemblies are available. Refer to Conax
Technologies’ Catalog 5001C.
Tel: 1.716.684.4500 • USA Toll Free: 1.800.223.2389 • Email: info@conaxtechnologies.com • Website: www.conaxtechnologies.com
"