RH0 pages
TM
www.temprecord.com
Humidity & Temperature – Multi Use
➤ urable, Reliable & Tamper Proof – Temperature and
D
Humidity recording where product life and product integrity
are critical!
➤ pplications – Monitoring of RH and/or temperature sensitive
A
products or equipment in storage, laboratories, production
processes or transit. Uses include analysis of weight loss due
to inadequate RH environment. Weight loss during storage
means loss in product value. Applications also include
monitoring of process/manufacture environments, HVAC
systems within multi-level buildings or labs.
➤ 2K non-volatile memory (16K Temperature and 16K
3
Relative Humidity)
➤ alibrated temperature range is accurate to 0.2°C / 0.35°F
C
and Relative Humidity to 3% RH (Logger Case temperature
range is -25°C to +60°C or -13°F to +140°F)
➤ -Point Temperature and 2-Point RH Calibration built
3
into logger:
• superior accuracy guarantee
• no need for corrections or complex calculations to be
applied to downloaded data
• voids human data entry errors, providing data integrity
a
and peace of mind
➤ alibrated and corrected at 5-Points to ISO/IEC 17025
C
Calibration Laboratory Standards
• traceable Calibration Certificate included
➤ emperature Units: °C or °F and % RH
T
➤ isplay Resolution: 0.01°C or 0.01°F or 0.1% RH
D
➤ TRW software:
• Hi-speed download
• Serial RS-232 or USB connectivity
• compatible with Windows XP, Vista and Windows 7
➤ ompliance made easy with automated PDF report
C
generation. Use speed-buttons from TRW software
to:
• save as *.TR and/or *.PDF files
• export to MS Excel – part or all samples
• email
• print
➤ ave time with Auto Mode (for batch
S
programming):
• Start / Stop / Status display
• download/read recorded data
• save downloaded file as *.TR or *.PDF
• print file
• email file
• reuse logger
➤ Fully welded food grade plastic case
• FDA approved
• recyclable
➤ Case protection rating:
splashproof to IP56 or
NEMA 4X