Gasmet Cr-40000 pages
v1.9
CR4000 FTIR Gas Analyzer
Multicomponent FTIR Gas Analyzer
Gasmet In-lab Series includes quantitative multicomponent
gas analyzers for laboratory research applications. The
Gasmet CR4000 incorporates a Fourier transform infrared,
FTIR spectrometer, a temperature controlled sample cell, and
signal processing electronics. Liquid nitrogen cooled MCT
detector has the highest performance available.
Spectrometer
Resolution:
Recommended 8 cm-1 or 4 cm-1
Scan frequency:
10 scans / s or faster
Detector:
Liquid N2 cooled MCT
The Gasmet CR4000 is designed for highest sensitivity and
speed. It is an ideal tool for example for emissions
measurements, catalyst research applications, or to measure
any compounds in wet corrosive gas mixtures. The longer path
lengths of the sample cell provide low detection limits. The
gold coated sample cell can be heated up to 180 ºC.
Source:
SiC, 1550 K
Optics material:
ZnSe (beamsplitter and windows)
Wave number range:
600 - 4200 cm-1
Sample cell
Mirrors:
Fixed, protected gold coating
0.4 liters
Connectors:
General parameters
Multi-pass, fixed path length 2.5 m
100 % rhodium coated aluminium
Volume:
CR-models scanning and analysis speed can be configured
according to customers need.
Structure:
Material:
Inlet Swagelok 6 mm
Outlet Swagelok 8 mm
Gaskets:
Viton® O-rings
Temperature:
180 °C, maximum
Window material:
ZnSe
Measuring principle:
Fourier transform infrared, FTIR
Performance:
Simultaneous analysis of up to 50
gas compounds
Response time, T90:
Typically < 120 s, depending on
the gas flow and measurement
time
Operating temperature:
5 - 30°C, non-condensing
Measuring parameters
Storage temperature:
-20 - 60°C, non-condensing
Zero point calibration:
Power supply:
100-115 or 230 V / 50 -60 Hz
24 hours, calibration with nitrogen
(4.0 or higher N2 recommended)
Power consumption:
300 W
Zero point drift:
< 2 % of measuring range per zero
"