VM6100 PAPERLESS RECORDER0 pages
Specifications Sheet
No.SPEC-1137E
(1st Edition)
VM6100
PAPERLESS RECORDER
OUTLINE
This is a paperless recorder that displays measured data
on the LCD in real time and stores data in CompactFlash.
The type of input such as thermocouple, resistance bulb,
D.C. voltage (current), etc. can be arbitrarily set to 18 channels at the maximum.
The data stored in CompactFlash can be regenerated on
the screen, and the use of supplied support software allows
the data to be regenerated on a PC screen.
The data recorded in ASCII format can be directly read in
a spreadsheet such as Excel, which facilitates the processing
on a PC. (The data recorded in binary format cannot be
read in.)
FEATURES
1. Large capacity storage by CompactFlash
Measured data is periodically stored in CompactFlash.
In case of 256 MB, for example, display files for about
one year and a half (display refresh cycle 30 sec) can
be taken up (in case of ASCII data format, 9 channels,
maximum/minimum recording).
2. Quick search and display of past data
Data stored in CompactFlash can be displayed in
succession by scrolling the screen.
3. Various display capability
Depending on the object of measurement, the most
suitable display format can be selected from a variety
of formats including bar graph display, trend display,
digital display,etc.
4. PC support software supplied as standard
Loader software that enables easy display and change
of set data and data viewer software that regenerates
the data stored in CompactFlash are supplied as
standard.
5. Compact size
160 (W) × 144 (H) × 185 (D) mm(Panel mounting),
Compact and as light as about 1.5 kg (9-point input,
without option).
6. 18-point recording (Option)
12 types of thermocouples, 2 types of resistance
bulbs and DC voltage/current input can be recorded
up to 18 points.
7. Communication function
RS-485 MODBUS communication is available.
8. LCD extinguishing function
Automatically extinguishes the LCD if nothing is
operated for certain time. You can set the time after a
lapse of which the LCD is extinguished via parameter
"LCD extinguishing time". The settable range is 0 to 60
minutes.
Setting at 0 minute overrides the function, whereby the
LCD will never extinguish.
This function prevents the backlight life from shortening
uselessly. During the extinguishment, the power consumption can be reduced.
9. Ethernet function(Option)
FTP,Web server, e-mail and MODBUS-TCP are available
using 10Base-T.
SPECIFICATIONS
Input system
Number of input points:9 points or 18 points (Can be
selected at the time of purchase)
Input circuit: Input mutual isolation
Resistance bulb measured current:
about. 1 mA
Measuring cycles:9 or 18 points….100ms cycles
Input types: Thermocouple, resistance bulb, DC voltage,
and DC current (Shunt resistorsvolt are
fitted in input terminals).
Note) Provide a shunt resistor (type:
WPSR222B000001A) separately.
Measuring range
Input types
Thermocouple
Resistance bulb
DC voltage
Reference range
B
R
S
K
E
J
T
N
W
L
U
PN
JPt100
Pt100
50mV
500mV
1-5V
0-5V
400.0
0.0
0.0
-200.0
-200.0
-200.0
-200.0
0.0
0.0
-200.0
-200.0
0.0
-200.0
-200.0
0.00
0.0
1.000
0.000
to 1760.0°C
to 1760.0°C
to 1760.0°C
to 1370.0°C
to 800.0°C
to 1100.0°C
to 400.0°C
to 1300.0°C
to 1760.0°C
to 900.0°C
to 400.0°C
to 1300.0°C
to 600.0°C
to 600.0°C
to 50.00mV
to 500.0mV
to 5.000V
to 5.000V
Note) B, R, S, K, E, J, T, N : JIS C 1602, DIN IEC 584-1
W : 5%Re-26%Re • W (Hoskins Mfg. Co. USA)
L : Fe-Cu • Ni (DIN 43710)
U : Cu-Cu • Ni (DIN 43710)
PN: Platinel2
JPt100 : JIS C 1604-1989 (Old JIS Pt 100)
Pt100 : JIS 1604, DIN IEC 751