Safety Standards Reference Guide0 pages
COMMON SAFETY STANDARD REFERENCE CHART
Standard / Harmonized
Standard
Testing Type
IEC/UL 60601-1 3rd
Performance
Edition Medical Electrical
Equipment
Production*
IEC 61730-2 UL 1703
Photovoltaic Modules &
Panels
IEC 60335-1 Household
Electrical Appliances
UL 60335-1 Household
Electrical Appliances
IEC 60598-1 Luminaires
DIELECTRIC WITHSTAND
Test Voltage
500 – 4000 VAC or
707 – 5656 VDC
60 s
GROUND BOND/CONTINUITY
Test Current V Limit
Max. R
Test Time
10-25 A
≤6V
≤ 0.1 Ω
5s
1 or 60 s
10-25 A
≤6V
≤ 0.1 Ω
5s
2.5 x Max
Over
Current
Protection
≤ 12 V
≤ 0.1 Ω
120 s
Performance
1000 VAC + 2 x rated V or
2000 VAC + 4 x rated V
50 uA
60 s
Production
1000 VAC + 2 x rated V or
(1000 VDC + 2 x rated V) X 120%
50 uA
1 or 60 s
Performance
500 – 2400 VAC x rated V +
2400 VAC
No
Breakdown
60 s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
≤ 120 s
Production
400 – 2500 VAC
5-30 mA
1s
≥ 10 A
≤ 12 V
0.1 – 0.2 Ω
No time
specified
60 s
40 A
≤ 6.5 V
≤ 0.5 Ω
120 s
Performance
500V – 2400 VAC x rated V + 2400
No
VAC
Breakdown
Continuity
Production
400 – 2500 VAC
5-30 mA
1s
40 A
≤ 12 V
0.1 – 0.2 Ω
No time
specified
Performance
500 – 4 x rated V + 2000 VAC
No
Breakdown
60 s
≥ 10 A
≤ 12 V
≤ 0.5 Ω
60 s
≤ 0.1 Ω
120 s
Not Specified - Responsibility of Manufacturer
Performance 1000 VAC - 1000 VAC x 2 x rated V
Production
IEC/UL 61010-1 &
Performance
CSA 22.2 No. 61010-1
Laboratory Control Test &
Measurement Equipment Production
EN 60204-1 Electrical
Equipment of Machines
No
Breakdown
1000 – 3000 VAC
Production
UL 1598 Luminaires
Max I.
Test
Time
Performance
1200 VAC
840 - 11940 VAC or
1200 - 7500 VDC
No
Breakdown
2 x rated V or
1000 VAC
No
Breakdown
Production
Performance
UL 2202 Electric
Vehicle Charging System
Equipment
Production
60 s
1s
5 - 60 s
30 A
≤4V
Continuity
25 or 30 A
≤ 10 V or ≤ 0.1 Ω or <4 V
≤ 12 V
0.133 Ω
5 s max
ramp up
2 s dwell
1s
≤ 0.1 Ω
500 VAC or 1000 VAC +
2 x rated V
1000 – 1700 VAC + 3.4 x rated V
1200 VAC + rated V or DC
Equivalent
Performance
1000 VAC + 2 x rated V or DC
equivalent
Production
1000 - 3000 VAC
arisafety.com
No
Breakdown
No
Breakdown
Performance
UL 45A Portable
Electrical Appliances
Continuity
0.2 - 10 A
≤ 24 V
Refer to
No time
Section 18.2.2 specified
60 s
≤ 60 A
≤ 12 V
Continuity
60 or 1 s
Continuity
60 s
No
Breakdown
1000 – 3000 VAC or
1414 – 4242 VDC
No
Breakdown
UL 60950-1 CSA 22.2
Performance
No. 60950-1 Information
Technology Equipment
Production
1000 – 3000 VAC or
1414 – 4242 VDC
No
Breakdown
Continuity
60 s
Continuity
1s
Continuity
120 s
30 A
≤ 12 V
≤ 0.1 Ω
60 s
1-4s
25 A
≤ 12 V
≤ 0.1 Ω
1-4 s
60 s
≤ 40 A
≤ 12 V
≤ 0.1 Ω
60 s
1–6s
Continuity
*As a result of performing risk analysis, many medical device manufacturers are performing leakage tests as part of 100% production line testing.
SAFETY IS OUR ONLY FOCUS®
120 –
240 s
Not Specified - Responsibility of Manufacturer
EN 60950-1 EN 50116 Performance
Information Technology
Equipment
Production
30
60 or
120 s
Not Specified - Responsibility of Manufacturer
IEC 61851-1 Electric
Vehicle Conductive
Charging System
Production
No
Breakdown
arisafety.com
"